Skip to content

ZEVO™ Anterior Cervical Plate System G7703515Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW G7703515 ZEVO VAR SD 3.5MM X 15MM · Class 2 · Implant

All models of ZEVO Anterior Cervical Plate System (39)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169398481), received between Jun 7, 2019 and Aug 6, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K141632); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING PATIENT HAVING SPINAL THERAPY FOR CERVICAL SPONDYLOTIC MYELOPATHY. IT WAS REPORTED THAT DURING THE SURGERY; THE SCREW LOCKING ISSUE OCCURRED. THE 90° SET SCREW WAS ROTATED, BUT IT TURNED ONCE BECAME LOOSE AND DID NOT STOP. THE SCREW AND PLATE WERE REMOVED AND REINSERTED. THERE WAS A DELAY OF LESS THAN 60 MINUTES AND NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22724767 ↗
  • MalfunctionJun 7, 2019Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR CERVICAL FIXATION OF C3-C6 DUE TO CERVICAL SPONDYLOTIC MYELOPATHY. INTRA-OP, THE HEAD OF THE SCREW AT C4 BROKE DURING INSERTION. THE HEAD PART OF THE BROKEN SCREW WAS SUCCESSFULLY REMOVED FROM PATIENT'S BODY. ONLY THE SHAFT OF THE SCREW REMAINS IMPLANTED IN THE PATIENT BODY AS IT WAS VERY DIFFICULT TO REMOVE. NO INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8677868 ↗
  • MalfunctionJun 7, 2019Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR CERVICAL FIXATION OF C3-C6 DUE TO CERVICAL SPONDYLOTIC MYELOPATHY. INTRA-OP, THE HEAD OF THE SCREW AT C4 BROKE DURING INSERTION. THE HEAD PART OF THE BROKEN SCREW WAS SUCCESSFULLY REMOVED FROM PATIENT'S BODY. ONLY THE SHAFT OF THE SCREW REMAINS IMPLANTED IN THE PATIENT BODY AS IT WAS VERY DIFFICULT TO REMOVE. NO INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8677848 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141632. One submission often covers many models, so check the product and lot details in each record.

  • Z-1750-2018 · Mar 23, 2018 · Terminated
    Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates