CD HORIZON® Spinal System 25530019590Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW 25530019590 BS CNCMAS 9.5X90 TC S · Class 2 · Implant

All models of CD HORIZON Spinal System (342)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169414310), received between Mar 27, 2024 and Mar 27, 2024 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Foreign Body Reaction1
  • Post Operative Wound Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAD OLIF (L2/3/4) AND T10~SAI FIXATION AND L5/SPLIF FROM THE REAR. IT WAS REPORTED THAT THE LOOSE SCREW CAUSED INFECTION IN PATIENT. ALTHOUGH FIXATION WAS PERFORMED FROM T10 TO SAI, A WOUND APPEARED ON THE SKIN OF THE RIGHT SAI AREA AND PUS ALSO CAME OUT. WHEN IT WAS DEPLOYED, THE TISSUE FROM THE SKIN SURFACE TO THE SCREW HEAD CONTINUED TO HAVE DARK GRANULATION, AND THERE WAS PUS IN THE SURROUNDING AREA. THE SET SCREW IN THE SCREW HEAD WAS LOOSE ENOUGH THAT IT COULD BE REMOVED WITH FORCEPS, AND THE SURROUNDING TISSUE SHOWED DARK SPOTS AND BLACK POWDERY MATTER. THE SCREW THAT HAD BEEN INSERTED HAD A CLEAR ZONE ON THE IMAGE, BUT IT WAS REASONABLY EFFECTIVE. THE RIGHT SIDE WAS IN A NOTICEABLE CONDITION, BUT WHEN CHECKED THE LEFT SIDE, IT WAS IN THE SAME CONDITION. A REVISION SURGERY WAS PERFORMED DUE TO SUSPICION OF METALLOSIS DUE TO WEAR PARTICLES OF THE IMPLANT, BLACK GRANULATION ON THE SOFT TISSUES, WOUNDS ON THE WOUND SURFACE AND PUS IN THE VICINITY. IMPLANT WAS CUT AND REMOVED. INITIAL SURGERY WAS DONE ON (B)(6) 2023. ADDITIONAL INFORMATION RECEIVED. THE SET SCREWS WERE L

    Patient problems: Foreign Body Reaction; Post Operative Wound Infection

    MAUDE report 18984825 ↗