ELEVATE™ Spinal System 8657002Recalls & reported problems

Medtronic Sofamor Danek, Inc. · INSTR INNR LOCKING TUBE 8657002 ELEVATE · Class 2

All models of ELEVATE Spinal System (33)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00643169432185), received between Aug 20, 2020 and Nov 26, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Material Deformation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 26, 2025Break; Break

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL FUSION THERAPY. IT WAS REPORTED THAT BOTH THE INNER SLEEVES COMING OUT, AND THE DISTAL THREADS OF THE ELEVATE INNER SLEEVE BREAKING OFF. IT WAS CONFIRMED THAT NO FRAGMENT OF THE INSTRUMENT REMAINED IN THE PATIENT. THERE WAS NO PATIENT INVOLVED AND NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23654354 ↗
  • MalfunctionApr 5, 2022Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING INSTRUMENTS USED FOR SPINAL SURGERY. PROCEDURE PERFORMED IN THE EVENT WAS TLIF. PRE-OP DIAGNOSIS WAS STENOSIS. IT WAS REPORTED THAT INTRA-OPERATIVELY, THREADS OF THE INSERTER INNER LOCKING TUBE WERE BROKEN. THE PRODUCT CAME IN CONTACT WITH PATIENT. THERE WAS NO DELAY IN THE OVERALL PROCEDURE. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14015310 ↗
  • MalfunctionSep 24, 2021Break; Material Deformation; Break; Material Deformation

    THE INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA A MANUFACTURING REPRESENTATIVE REGARDING A PATIENT FOR SPINAL THERAPY. IT WAS REPORTED THAT THE BROKEN SERRATION OF CAGE INSERTER FOR ELEVATE CAGE. INABILITY TO ENGAGE PROPERLY ON IMPLANT. THERE WAS NO PATIENT SYMPTOMS REPORTED IN THIS EVENT. ADDITIONAL INFORMATION RECEIVED STATES THAT ELEVATE INSERTER WAS STRIPPED AT TIME OF INSERTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12524440 ↗
  • MalfunctionApr 1, 2021Break; Break

    MEDTRONIC RECEIVED INFORMATION REGARDING NAVIGATION AND SPINAL INSTRUMENTS BEING USED INTRA-OPERATIVELY OF A SPINAL PROCEDURE. IT WAS REPORTED THAT INSTRUMENTS BROKE DUE TO HARD PATIENT BONES. THERE WAS NO DELAY AND NO IMPACT ON THE PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11605763 ↗
  • MalfunctionApr 1, 2021Break; Break

    MEDTRONIC RECEIVED INFORMATION REGARDING NAVIGATION AND SPINAL INSTRUMENTS BEING USED INTRA-OPERATIVELY OF A SPINAL PROCEDURE. IT WAS REPORTED THAT INSTRUMENTS BROKE DUE TO HARD PATIENT BONES. THERE WAS NO DELAY AND NO IMPACT ON THE PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11605700 ↗
  • MalfunctionNov 26, 2020Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING AN EVENT THAT OCCURRED DURING AN L3-4-5-1 TLIF PROCEDURE IN A PATIENT DIAGNOSED WITH SPINAL CORD STENOSIS. IT WAS REPORTED THAT THE THREADS WERE BROKEN. PRODUCT WAS RETURNED, AND IT WILL BE REPLACED WITH A MEDTRONIC PRODUCT. NO PATIENT INJURY / COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10910526 ↗
  • MalfunctionOct 14, 2020Break; Material Deformation; Break; Material Deformation

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OPERATIVE DIAGNOSIS FOR SPINAL STENOSIS PROCEDURE. IT WAS REPORTED THAT THE SURGEON NOTICED SHAVINGS FROM THREADED SHAFT WHEN ATTACHING IMPLANT. DURING IMPLANTATION BOTH TABS BROKE OFF ON INSERTER. THERE WAS NO FRAGMENT LEFT, OR CONTACT WITH THE PATIENT. THERE WERE NO SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10675900 ↗
  • MalfunctionAug 20, 2020Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OPERATIVE DIAGNOSIS OF L3/4;L4-S1 PLIF. IT WAS REPORTED THAT INTRA-OP, INSERTER ELEVATE AND ITS INNER LOCKING TUBE BREAK. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT. PRODUCTS WILL BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10429933 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →