ELEVATE™ Spinal System 8657005Recalls & reported problems
Medtronic Sofamor Danek, Inc. · DISTRACTOR 1 8657005 ELEVATE · Class 2
All models of ELEVATE Spinal System (33)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169432215), received between Aug 28, 2017 and Apr 7, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Known Impact Or Consequence To Patient1
- Spinal Cord Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 7, 2021Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT UNDERGOING A PROCEDURE OF TRANSFORAMINAL LUMBAR INTERBODY FUSION AT THE LEVELS OF L4-L5 WITH PRE-OPERATIVE DIAGNOSIS OF SPONDYLOLISTHESIS. IT WAS REPORTED THAT, THE INSTRUMENT WAS BROKEN DUE TO FATIGUE FAILURE WHILE IN DISC SPACE. WHILE REMOVING INSTRUMENT, THE DURA WAS TORN (DURAL TEAR). AN ADDITIONAL TREATMENT WAS PERFORMED AS A RESULT OF THIS EVENT. NO FRAGMENTS OF THE BROKEN INSTRUMENT REMAINED IN PATIENT. THERE WAS A DELAY OF 20 MINUTES OCCURRED TO REPAIR THE DURA. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: Spinal Cord Injury
MAUDE report 11626176 ↗MalfunctionAug 28, 2017Break; Break
PROCEDURE: TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF). IT WAS REPORTED THAT THE PATIENT UNDERWENT SPINAL SURGERY. THE DISTRACTOR USED DURING THE SURGERY COLLAPSED TO THE NEUTRAL POSITION AND BROKE INSIDE PATIENT'S DISC SPACE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. NO PART OF THE INSTRUMENT WAS LEFT INSIDE THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6823332 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page