Skip to content

CD HORIZON® Spinal System 54840047535Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW 54840047535 4.75 ATS MAS 7.5X35 · Class 2 · Implant

All models of CD HORIZON Spinal System (342)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169438668), received between Apr 24, 2023 and Apr 24, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 24, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH SCREW AND ROD USED IN L4-S1 TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF). IT WAS REPORTED THAT ROD ALMOST FULLY RELEASED FROM THE LEFT S1 SCREW HEAD WHILE CAP REMAINED ON. THE SCREWHEAD LOOKED LIKE FAULTY. THERE WAS SCREW HEAD MOVEMENT AND ROD MIGRATED OUT OF SCREW HEAD. PRE-OPERATIVE DIAGNOSIS FOR THE PATIENT WAS LEG PAIN. S1 SCREW AND ROD ON LEFT SIDE WERE EXPLANTED AS AN ADDITIONAL SURGERY ON 13 APRIL AND WERE REPLACED. PATIENT WAS EXPERIENCING LEG PAIN. THERE WERE NO FURTHER COMPLICATIONS THAT WERE REPORTED.

    Patient problems: Pain

    MAUDE report 16795321 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates