Skip to content

Delta® 42833Recalls & reported problems

Medtronic Ps Medical, Inc. · VALVE 42833DELTA BURR HOLE 12MM LEV 1.5 · Class 2 · Implant

All models of Delta (31)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169466838), received between Mar 2, 2021 and Mar 2, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Swelling/ Edema1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 2, 2021Break; Break

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE ETIOLOGY OF THE REOPERATION WAS THE SHUNT MALFUNCTION DUE TO EXTENSIVE INSERTION OF THE VENTRICULAR CATHETER INTO THE BRAIN PARENCHYMA. DURING THE INSPECTION OF THE CRANIAL FLAP, THERE WAS SWELLING UNDERNEATH. THE PHYSICIAN OPENED THE FLAP, AND THERE WAS CEREBROSPINAL FLUID AROUND THE VALVE MECHANISM. UNDER THE SURGICAL MICROSCOPE, IT WAS SEEN A SMALL RUPTURE AROUND THE VALVE MECHANISM, NOT IN THE RESERVOIR.

    Patient problems: Swelling/ Edema

    MAUDE report 11400060 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page