Skip to content

Endurant® II ETLW1624C82E (UDI 00643169467477)Recalls & reported problems

Medtronic, Inc. · STENT GRAFT ETLW1624C82E ENDUR II LIMB · Class 3 · Implant

All models of Endurant II (114)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169467477), received between Apr 12, 2018 and Oct 8, 2020 — classified as 2 injuries.

The most frequently reported specific device problem is “Failure To Adhere Or Bond” (1 report).

6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Failure To Adhere Or Bond1

Patient problems

  • Injury1
  • Numbness1
  • Obstruction/Occlusion1
  • Occlusion1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AN ENDURANT II BIFURCATE STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN 70 MM ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL COMPLAINING OF PROGRESSING LEG PAIN AND NUMBNESS. A CT WAS PERFORMED APPROXIMATELY 2 YEARS POST INDEX PROCEDURE, WHERE REVEALED A THROMBUS WITH COMPLETE OCCLUSION OF THE LEFT ENDURANT II LIMB AND A FOCAL THROMBUS IN THE RIGHT COMMON ILIAC ARTERY DISTAL TO THE ENDURANT II LIMB AND AT THE BIFURCATION OF THE COMMON ILIAC ARTERY. THE PHYSICIAN ATTEMPTED TO PERFORM A THROMBECTOMY OF THE LEFT ENDURANT II LIMB GRAFT AND A THROMBECTOMY TO REMOVE THE FOCAL THROMBUS IN THE RIGHT COMMON ILIAC WHICH BOTH ATTEMPTS WERE UNSUCCESSFUL. THE PHYSICIAN THEN DECIDED TO EXTEND THE ENDURANT II LIMB INTO THE RIGHT EXTERNAL ILIAC ARTERY AND COVER THE RIGHT INTERNAL ILIAC. A FEMORAL TO FEMORAL BYPASS WAS PERFORMED. IT WAS STATED THAT THE CT SHOWED NO EVIDENCE OF KINKING OR DISRUPTIONS OF THE ENDURANT II LIMBS. THE LUMEN OF THE LIMB GRAFTS MEASURED 15 TO 16 MM PROXIMALLY AND 23 TO 24 MM DISTALLY. AS PER THE PHYSICIAN, CAUSE OF THE EVENT CANNOT BE DETERMINED. NO ADDITIONAL CLINICAL SEQUALAE WERE REPORTED AND THE

    Patient problems: Numbness; Obstruction/Occlusion; Occlusion; Pain

    MAUDE report 10648963 ↗
  • InjuryApr 12, 2018Failure To Adhere Or Bond; Failure To Adhere Or Bond

    AN ENDURANT STENT GRAFT WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 31 MM AORTO ILIAC ANEURYSM. IT WAS REPORTED THAT THE PATIENT HAD AN ENLARGED LEFT COMMON ILIAC ARTERY AND LOSS OF SEAL IN THE LEFT COMMON ILIAC ARTERY. DURING THE REINTERVENTION PROCEDURE THE LEFT INTERNAL ILIAC ARTERY WAS EMBOLISED AND AN ENDURANT LIMB ETLW1610C124E WAS EXTENDED INTO THE EXTERNAL ILIAC ARTERY. THE EVENT HAS BEEN RESOLVED THROUGH SUCCESSFUL EXCLUSION OF THE ILIAC ANEURYSM. THE PHYSICIAN ATTRIBUTED THE CAUSE OF THE EVENT TO DISEASE PROGRESSION CAUSING LOSS OF SEAL. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

    Patient problems: Injury

    MAUDE report 7424571 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0290-2022 · Oct 15, 2021 · Open, Classified
    Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
  • Z-0292-2022 · Oct 15, 2021 · Open, Classified
    Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
  • Z-0291-2022 · Oct 15, 2021 · Open, Classified
    Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
  • Z-2136-2021 · Jun 7, 2021 · Open, Classified
    Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoili
  • Z-2135-2021 · Jun 7, 2021 · Open, Classified
    Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infra
  • Z-1543-2017 · Feb 27, 2017 · Terminated
    Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page