Endurant® II ETLW1613C82E (UDI 00643169467699)Recalls & reported problems
Medtronic, Inc. · STENT GRAFT ETLW1613C82E ENDUR II LIMB · Class 3 · Implant
All models of Endurant II (114)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 2
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00643169467699), received between Mar 7, 2017 and Jul 18, 2024 — classified as 2 deaths, 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Occlusion Within Device” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Occlusion Within Device2
- Device Expiration Issue1
- Leak/Splash1
Patient problems
- Occlusion2
- Pain2
- Hemorrhage/Blood Loss/Bleeding1
- Injury1
- No Consequences Or Impact To Patient1
- Obstruction/Occlusion1
- Ruptured Aneurysm1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
DeathJul 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN ENDURANT IIS BIFURCATE STENT GRAFT SYSTEM WAS IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A 66MM AAA. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, A LEFT SIDED OCCLUSION OCCURRED AFFECTING THE STENT GRAFTS ETLW1613C93E (B)(6) AND ETLW1613C82E (B)(6). INTERVENTION WAS PERFORMED AND A 32 ENDURANT AUI AND 36 ENDURANT CUFF WERE IMPLANTED. A FEM-FEM BYPASS WAS PERFORMED ALSO PERFORMED WITH THE IMPLANT OF A NON MDT PLUG. IT WAS ALSO NOTED THAT THERE WAS AN EXTRAVASATION IN THE PROXIMAL RIGHT EXTERNAL ILIAC ARTERY (REIA), DISTAL TO THE RIGHT LIMB CAUSED BY A NON-MEDTRONIC THROMBECTOMY CATHETER. THE REMOVAL OR DELIVERY OF THE STENT GRAFTS DID NOT CAUSE ANY INJURY IN THE REIA. PER THE PHYSICIAN THE PROBABLE CAUSE OF OCCLUSION WAS PATIENT ANATOMY. IT WAS NOTED THE PATIENT HAD CALCIFIED ACCESS VESSELS. THE EXTRAVASATION OF BLOOD WAS CAUSED BY THE THROMBECTOMY DEVICE. IT WAS THEN REPORTED THE PATIENT EXPIRED THE FOLLOWING DAY. THE CAUSE OF DEATH IS UNKNOWN BUT AS PER THE PHYSICIAN NOT RELATED TO THE ENDURANT STENT GRAFTS. NO ADDITIONAL CLINICAL SEQUELAE WERE PROVIDED .
Patient problems: Hemorrhage/Blood Loss/Bleeding; Obstruction/Occlusion
MAUDE report 19772427 ↗DeathNov 24, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN ENDURANT IIS STENT GRAFT SYSTEM AND VALIANT STENT GRAFT WAS IMPLANTED IN A PATIENT DURING AN UNKNOWN AORTIC ENDOVASCULAR PROCEDURE. IT WAS REPORTED APPROXIMATELY 1 YEAR POST THE INDEX PROCEDURE, THE PATIENT PRESENTED TO THE ER. A CT SCAN IDENTIFIED A TYPE IA ENDOLEAK AND RUPTURED AAA. THE PATIENT EXPIRED ON THE DAY OF THE RUPTURE. PER THE PHYSICIAN THE CAUSE OF THE ENDOLEAK RUPTURE AND DEATH IS UNDETERMINED.
Patient problems: Ruptured Aneurysm
MAUDE report 10894951 ↗InjuryJan 3, 2018Occlusion Within Device; Occlusion Within Device
AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE PATIENT HAD A VERY TORTUOUS ILIAC WITH A 90 DEGREE BEND. BECAUSE OF THIS, THE PHYSICIAN DECIDED TO EXTEND THE BIFURCATED STENT GRAFT DURING THE INDEX PROCEDURE, USING A LIMB. IT WAS REPORTED THAT DURING A FOLLOW-UP, IT WAS NOTED THAT THE PATIENT HAD AN OCCLUDED LEFT LIMB. THE PHYSICIAN TREATED THE OCCLUSION BY IMPLANTATION OF A STAINLESS-STEEL STENT AND REMOVAL OF THE CLOT. THE PHYSICIAN STATED THE CAUSE OF THE EVENT WAS ANATOMY RELATED, DUE TO THE TORTUOUS ILIAC. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: Occlusion
MAUDE report 7159521 ↗MalfunctionSep 22, 2017Device Expiration Issue; Device Expiration Issue
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE EMERGENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE HOSPITAL OWNED ENDURANT STENT GRAFT THAT WAS IMPLANTED ONE DAY POST EXPIRY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6885870 ↗InjuryAug 24, 2017Leak/Splash; Leak/Splash
AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF A GREATER THAN 55 MM IN DIAMETER ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT, APPROXIMATELY ONE YEAR AND SEVEN MONTHS POST INDEX PROCEDURE, THE PATIENT PRESENTED EMERGENTLY WITH ABDOMINAL AND BACK PAIN. IT WAS DISCOVERED THAT THE ENDURANT II STENT GRAFT SYSTEM HAD BILATERAL DISTAL TYPE I ENDOLEAKS. THE PHYSICIAN ELECTED TO IMPLANT BILATERAL ILIAC LIMB EXTENSIONS AND IMPLANTED A PROXIMAL EXTENSION. THE PROXIMAL EXTENSION WAS IMPLANTED AS A PROACTIVE THERAPY TO ELIMINATE ANY UNDEFINED OR NON-IDENTIFIED PROXIMAL ENDOLEAKS. THE EVENT WAS RESOLVED. PER THE PHYSICIAN, THE CAUSE OF THE EVENT WAS DUE TO DILATATION OF THE COMMON ILIAC ARTERIES FROM THE DISEASE STATE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: Injury; Pain
MAUDE report 6816118 ↗InjuryMar 7, 2017Occlusion Within Device; Occlusion Within Device
ADDITIONAL INFORMATION RECEIVED. THE OCCLUSION EXTENDED TO THE LEVEL OF THE FLOW DIVIDER AND KINKING WAS OBSERVED IN THE PROXIMAL IPSILATERAL LIMB OF THE AORTIC BIFURCATION.
Patient problems: Occlusion; Pain
MAUDE report 6385769 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.
- Z-0290-2022 · Oct 15, 2021 · Open, ClassifiedEndurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
- Z-0292-2022 · Oct 15, 2021 · Open, ClassifiedEndurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
- Z-0291-2022 · Oct 15, 2021 · Open, ClassifiedEndurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
- Z-2136-2021 · Jun 7, 2021 · Open, ClassifiedMedtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoili
- Z-2135-2021 · Jun 7, 2021 · Open, ClassifiedEndurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infra
- Z-1543-2017 · Feb 27, 2017 · TerminatedEndurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page