Medtronic Navigation, Inc. · STYLET 9735428 2 COIL SINGLE PACKAGE · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00643169490468), received between Apr 9, 2019 and Feb 26, 2026 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem3
  • Biocompatibility2
  • Contamination1
  • Device Sensing Problem1
  • Output Problem1
  • Use of Device Problem1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • No Clinical Signs, Symptoms or Conditions3
  • Unspecified Infection2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2026Contamination

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED IN AN UNKNOWN CONTEXT. IT WAS REPORTED THAT THE AXIEM WAS NOT STERILE. ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE ISSUE OCCURRED PRIOR TO A CATHETER PLACEMENT PROCEDURE. IT WAS CLARIFIED THAT THE STYLET CAME WITHOUT THE STERILITY SEAL CLOSED ON THE PACKAGING, NOR WAS IT MANIPULATED. THE PROCEDURE WAS COMPLETED USING ANOTHER CONSUMABLE. THERE WAS NO PATIENT PRESENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24455759 ↗
  • MalfunctionMar 12, 2024Device Sensing Problem; Device Sensing Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING AN UNKNOWN PROCEDURE. IT WAS REPORTED THAT THE STYLET WAS NOT RECOGNIZED BY NAVIGATION. THERE WAS A TRACKING ISSUE EXPERIENCED WITH THE STYLET. THE SYSTEM DID NOT RECOGNIZE THE NAME OF THE INSTRUMENT IN THE SOFTWARE. A DIFFERENT STYLET WAS NEEDED TO BE USED. THE SUSPECTED CAUSE OF THE EVENT WAS UNKNOWN. NO FURTHER INFORMATION WAS ABLE TO BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18888769 ↗
  • MalfunctionJul 26, 2022Use of Device Problem; Use of Device Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT WHEN PREPARING FOR A CASE THE SITE OPENED AN EM STYLET AND IT WAS BENT IN THE PACKAGING. SITE OPENED ANOTHER STYLET. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15103497 ↗
  • InjuryNov 15, 2021Biocompatibility; Biocompatibility

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED DURING A PROCEDURE. IT WAS REPORTED THAT AN INFECTION OCCURRED DURING THE PROCEDURE UTILIZING THE NAVIGATION STYLET. THERE WAS NO REPORTED DELAY TO THE PROCEDURE DUE TO THIS ISSUE.

    Patient problems: Unspecified Infection

    MAUDE report 12814769 ↗
  • InjuryNov 15, 2021Biocompatibility; Biocompatibility

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED DURING A PROCEDURE. IT WAS REPORTED THAT AN INFECTION OCCURRED DURING THE PROCEDURE UTILIZING THE NAVIGATION STYLET. THERE WAS NO REPORTED DELAY TO THE PROCEDURE DUE TO THIS ISSUE.

    Patient problems: Unspecified Infection

    MAUDE report 12814757 ↗
  • MalfunctionNov 5, 2019Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED INTRA-OPERATIVELY OF A CATHETER PLACEMENT PROCEDURE. IT WAS REPORTED THAT THEY COULD NOT SEE THE STYLETS. THEY USED ONE STYLET THAT DIDN'T DETECT AT ALL. THEY OPENED A SECOND STYLET WHICH DETECTED FOR 3 PASSES, THEN PICKING IT UP AGAIN, IT NO LONGER DETECTED. THEY OPENED A THIRD STYLET THAT DETECTED FOR 3 PASSES THEN STOPPED DETECTING. THE ISSUE EXTENDED THE SURGICAL TIME BY LESS THAN ONE HOUR. THERE WAS NO IMPACT ON THE PATIENT OUTCOME. THE SURGERY WAS COMPLETED WITH THE USE OF A C-ARM. THE TRACKING ISSUE WAS THOUGHT TO BE RELATED TO THE STYLETS AS THE SYSTEM NEVER LOST THE PATIENT TRACKER AND THE STYLET WAS GREEN ON THE INSTRUMENT SCREEN, WHICH INDICATED IT WAS PLUGGED INTO THE EMITTER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9282248 ↗
  • MalfunctionNov 5, 2019Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED INTRA-OPERATIVELY OF A CATHETER PLACEMENT PROCEDURE. IT WAS REPORTED THAT THEY COULD NOT SEE THE STYLETS. THEY USED ONE STYLET THAT DIDN'T DETECT AT ALL. THEY OPENED A SECOND STYLET WHICH DETECTED FOR 3 PASSES, THEN PICKING IT UP AGAIN, IT NO LONGER DETECTED. THEY OPENED A THIRD STYLET THAT DETECTED FOR 3 PASSES THEN STOPPED DETECTING. THE ISSUE EXTENDED THE SURGICAL TIME BY LESS THAN ONE HOUR. THERE WAS NO IMPACT ON THE PATIENT OUTCOME. THE SURGERY WAS COMPLETED WITH THE USE OF A C-ARM. THE TRACKING ISSUE WAS THOUGHT TO BE RELATED TO THE STYLETS AS THE SYSTEM NEVER LOST THE PATIENT TRACKER AND THE STYLET WAS GREEN ON THE INSTRUMENT SCREEN, WHICH INDICATED IT WAS PLUGGED INTO THE EMITTER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9282140 ↗
  • MalfunctionNov 5, 2019Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED INTRA-OPERATIVELY OF A CATHETER PLACEMENT PROCEDURE. IT WAS REPORTED THAT THEY COULD NOT SEE THE STYLETS. THEY USED ONE STYLET THAT DIDN'T DETECT AT ALL. THEY OPENED A SECOND STYLET WHICH DETECTED FOR 3 PASSES, THEN PICKING IT UP AGAIN, IT NO LONGER DETECTED. THEY OPENED A THIRD STYLET THAT DETECTED FOR 3 PASSES THEN STOPPED DETECTING. THE ISSUE EXTENDED THE SURGICAL TIME BY LESS THAN ONE HOUR. THERE WAS NO IMPACT ON THE PATIENT OUTCOME. THE SURGERY WAS COMPLETED WITH THE USE OF A C-ARM. THE TRACKING ISSUE WAS THOUGHT TO BE RELATED TO THE STYLETS AS THE SYSTEM NEVER LOST THE PATIENT TRACKER AND THE STYLET WAS GREEN ON THE INSTRUMENT SCREEN, WHICH INDICATED IT WAS PLUGGED INTO THE EMITTER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9282123 ↗
  • MalfunctionApr 9, 2019Output Problem; Output Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED FOR A CATHETER PLACEMENT PROCEDURE. THE ISSUE OCCURRED INTRAOPERATIVELY AND CAUSED NO DELAY TO THE SURGERY. IT WAS REPORTED THAT THERE WAS NO PROBLEM TILL REGISTRATION. THE OPERATIVE FIELD BECAME STERILE AND WHEN THE STYLET GOT CLOSE TO THE STERILE OPERATIVE FIELD, THE RECOGNITION WAS UNSTABLE. ADJUSTING THE POSITION OF METAL AND EMITTER WAS PERFORMED, BUT NO IMPROVEMENT WAS SEEN. THE PROCEDURE WAS COMPLETED WITH THE CONTINUOUS USE OF THE NAVIGATION SYSTEM AND THERE WAS NO IMPACT ON PATIENT OUTCOME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8495824 ↗

Other Neurological stereotaxic Instrument devices