REVEAL® XT 9529 (UDI 00643169494893)Recalls & reported problems
Medtronic, Inc. · MONITOR 9529 REVEAL XT USA · Class 2 · Implant
Quick summary
The Reveal XT is a small heart monitor inserted under the skin to watch for rhythm problems. The 51 reports mostly describe the monitor misreading the heartbeat — picking up too little or too much signal, which led it to flag false slow-heart and fast-heart episodes — and a remote unit that failed to send data to the clinic. There are50 malfunction reports, 1 injury involving an infection that would not heal at the implant site, and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 51
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 51 adverse event reports that list this exact device (DI 00643169494893), received between Jun 8, 2017 and Apr 24, 2023 — classified as 1 injury, 50 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (28 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Under-Sensing28
- Over-Sensing16
- Invalid Sensing4
- Signal Artifact/Noise4
- Communication or Transmission Problem3
- Device Sensing Problem3
- Failure to Interrogate3
- Premature Discharge of Battery3
Patient problems
- No Known Impact Or Consequence To Patient38
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient3
- Impaired Healing1
- No Patient Involvement1
- Unspecified Infection1
Recent adverse event reports
The 25 most recent of 51 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 24, 2023Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE REMOTE MONITOR WAS NOT COLLECTING INFORMATION FROM THE IMPLANTABLE CARDIAC MONITOR (ICM) DEVICE. THE ICM HAD NO TELEMETRY. TROUBLESHOOTING STEPS WERE TAKEN AND STILL NOT GETTING THE PROGRESS BAR. THE CALLER WAS REFERRED TO CLINIC. THE REMOTE MONITOR REMAINS IN USE AND THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16794804 ↗MalfunctionApr 7, 2023Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE REMOTE MONITOR WAS UNABLE TO ESTABLISH TELEMETRY WITH THE IMPLANTABLE CARDIAC MONITOR (ICM). IT WAS ALSO RE PORTED THAT AN ERROR CODE OF 7332 APPEARED INDICATING THE CONNECTION TO LOOK UP THE AFFILIATED DEVICE IN THE REMOTE MONITORING NETWORK WAS UNSUCCESSFUL OR TIMED OUT. TROUBLESHOOTING STEPS WERE EXHAUSTED AND STILL PROGRESS BAR WOULD NOT START. THE PATIENT WAS REFER RED TO THE CLINIC AND TO TAKE THE MONITOR TO THE APPOINTMENT. THE MONITOR REMAINS IN USE. THE ICM REMAINS IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16700184 ↗MalfunctionNov 23, 2022Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE REMOTE MONITOR WAS UNABLE TO ESTABLISH TELEMETRY WITH THE IMPLANTABLE CARDIAC MONITOR (ICM). TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE ICM REMAINS IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15850537 ↗MalfunctionAug 31, 2021Signal Artifact/Noise; Over-Sensing; Under-Sensing; Signal Artifact/Noise; Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED UNDERSENSING ON ASYSTOLE AND BRADYCARDIA EPISODES. IT WAS FURTHER REPORTED THAT THE DEVICE EXPERIENCED OVERSENSING NOISE ON EPISODES. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12390597 ↗MalfunctionMay 27, 2021Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) DETECTED FALSE PAUSE EPISODES DUE TO UNDERSENSING. IT WAS FURTHER REPORTED THAT THE ICM DETECTED FALSE VENTRICULAR TACHYCARDIA (VT) DUE TO OVERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11892837 ↗MalfunctionMay 7, 2021Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED UNDERSENSING R-WAVES/PREMATURE VENTRICULAR CONTRACTIONS (PVC) WHICH RESULTED IN FALSE PAUSE AND BRADYCARDIA EPISODES. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11788908 ↗InjuryMar 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) PATIENT EXPERIENCED AN INFECTION AND THE SITE WOULD NOT HEAL PROPERLY. THE DEVICE WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Impaired Healing; Unspecified Infection
MAUDE report 11409337 ↗MalfunctionMar 1, 2021Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) HAD NO TELEMETRY. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11391641 ↗MalfunctionNov 25, 2020Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED UNDERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10904411 ↗MalfunctionJun 23, 2020Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED NO TELEMETRY AS THE READER WAS REPOSITIONED MULTIPLE TIMES. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10183008 ↗MalfunctionJun 15, 2020Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) DETECTED FASCICULAR VENTRICULAR TACHYCARDIA (FVT) DUE TO OVERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10152403 ↗MalfunctionMay 20, 2020Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED T-WAVE OVERSENSING (TWOS). THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10073867 ↗MalfunctionFeb 28, 2020Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED OVERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9768267 ↗MalfunctionFeb 10, 2020Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXHIBITED OVERSENSING AND UNDERSENSING. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9687328 ↗MalfunctionSep 20, 2019Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED A PAUSE EPISODE THAT LOOKS TO BE LOSS OF CONTACT. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9096623 ↗MalfunctionSep 18, 2019Under-Sensing; Device Sensing Problem; Under-Sensing; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED A FALSE PAUSE EPISODE DUE TO UNDERSENSING OF THE QRS COMPLEX. IT WAS FURTHER REPORTED THAT TACHYCARDIA EPISODES DO NOT APPEAR TO BE VENTRICULAR TACHYCARDIA (VT).THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9080950 ↗MalfunctionSep 11, 2019Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) DETECTED A PAUSE EPISODE DUE TO UNDERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8988143 ↗MalfunctionAug 22, 2019Under-Sensing; Decreased Sensitivity; Under-Sensing; Decreased Sensitivity
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED DIMINISHED R WAVES AND LOSS OF CONTACT. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8920619 ↗MalfunctionAug 15, 2019Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED RIGHT VENTRICULAR (RV) UNDERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8896191 ↗MalfunctionJun 28, 2019Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) DETECTED FALSE PAUSE DUE TO UNDERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8743105 ↗MalfunctionMay 28, 2019Failure to Interrogate; Battery Problem; Failure to Interrogate; Battery Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) BATTERY DEPLETED PREMATURELY, AND COULD NOT BE INTERROGATED. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8645358 ↗MalfunctionApr 17, 2019Signal Artifact/Noise; Over-Sensing; Signal Artifact/Noise; Over-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED OVERSENSING ARTIFACT. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8522044 ↗MalfunctionApr 12, 2019Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED R-WAVE UNDERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8508270 ↗MalfunctionApr 10, 2019Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED OVERSENSING AND UNDERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8499662 ↗MalfunctionApr 5, 2019Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) EXPERIENCED VENTRICULAR UNDERSENSING. THE ICM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8486876 ↗
Other DETECTOR AND ALARM, ARRHYTHMIA devices
- Reveal LINQ™ LNQ11Medtronic, Inc. · DETECTOR AND ALARM, ARRHYTHMIA1 recall7,581 reports
- Confirm Rx™ DM3500St. Jude Medical, Inc. · DETECTOR AND ALARM, ARRHYTHMIA2 recalls2 deaths5,149 reports
- Reveal LINQ TM LNQ11Medtronic, Inc. · DETECTOR AND ALARM, ARRHYTHMIA1 death1,592 reports
- Jot Dx™ DM4500St. Jude Medical, Inc. · DETECTOR AND ALARM, ARRHYTHMIA1 death1,005 reports
- Reveal LINQ™ LNQ11Medtronic, Inc. · DETECTOR AND ALARM, ARRHYTHMIA1 death930 reports
- LINQ II™ LNQ22Medtronic, Inc. · DETECTOR AND ALARM, ARRHYTHMIA2 recalls673 reports
- Reveal LINQ™ LNQ11Medtronic, Inc. · DETECTOR AND ALARM, ARRHYTHMIA431 reports
- Reveal LINQ™ LNQ11Medtronic, Inc. · DETECTOR AND ALARM, ARRHYTHMIA1 recall343 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.