Skip to content

CD HORIZON® SOLERA® Spinal System 641100060Recalls & reported problems

Medtronic Sofamor Danek, Inc. · ROD 641100060 60MM PERC ROD 4.75 CCM ST · Class 2 · Implant

All models of CD HORIZON SOLERA Spinal System (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169501409), received between Sep 11, 2017 and Sep 11, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration or Expulsion of Device1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2017Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT THE PATIENT HAD UNDERGONE A SPINAL FUSION AT L1-L3 DUE TO BURST FRACTURE AT L1-3 OR L2 VERTEBRAL FRACTURE. POST-OP, WHEN BONE UNION WAS ACHIEVED, A REMOVAL SURGERY WAS CONDUCTED. DURING THIS REVISION SURGERY, THE IMPLANTED ROD IN THE LEFT SIDE OF L3 WAS FOUND TO BE DETACHED FROM SCREW HEAD ON THE CAUDAL SIDE. THE SET SCREW WAS CHECKED AND WAS FOUND TO BE BROKEN, BUT THIS SET SCREW REMAINED IN THE SCREW HEAD. ALL THE IMPLANTS WERE REMOVED IN THE REVISION SURGERY, WITH NO FRAGMENT OF THE BROKEN PRODUCT REMAINING INSIDE THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6854434 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates