Skip to content

Freezor® 307F3Recalls & reported problems

Medtronic, Inc. · CATHETER 307F3 FREEZOR3 MED BLUE 7F · Class 3

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169504752), received between Jun 9, 2023 and Jun 9, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Protective Measures Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P020045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Protective Measures Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 9, 2023Adverse Event Without Identified Device or Use Problem; Protective Measures Problem; Adverse Event Without Identified Device or Use Problem; Protective Measures Problem

    IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE INJECTION WAS STOPPED BECAUSE THE CRYOMAPPING TEMPERATURE EXCEEDED THE PRESET VALUE BY MORE THAN FIFTEEN DEGREES. THE CASE WAS ABORTED AND THE PATIENT WAS UNDER GENERAL ANESTHESIA. TEST ABLATIONS WERE PERFORMED AND THE CONSOLE APPEARED TO BE CONTROLLING PRESSURE AND FLOW IN THE EXPECTED RANGES. A PREVENTATIVE MAINTENANCE VISIT WAS RECOMMENDED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17097053 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2777-2015 · Sep 1, 2015 · Terminated
    Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.
  • Z-1452-06 · Aug 4, 2006 · Terminated
    7F Freezor Cardiac Cryoablation catheter, REF # 207F1. CryoCath Technologies Inc.

Other Cardiac ablation percutaneous catheter devices

All Cardiac ablation percutaneous catheter devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.