N/A 9734261 (UDI 00643169510531)Recalls & reported problems
Medtronic Navigation, Inc. · UPGRD PRGM 9734261 StealthStation S7 · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169510531), received between Jul 25, 2017 and Jul 25, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 26, 2018Terminated
An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.
Z-0552-2019 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Use of Device Problem1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2017Use of Device Problem; Device Operates Differently Than Expected; Use of Device Problem; Device Operates Differently Than Expected
A MEDTRONIC REPRESENTATIVE REPORTED THAT PRIOR TO A VISUALASE PROCEDURE, THE SURGEON MONITOR ON THE NAVIGATION SYSTEM BEGAN FLICKERING. THE SYSTEM WAS ON FOR A FEW HOURS WHEN THE ISSUE BEGAN. THE CABLE CONNECTIONS WERE CHECKED AND THE MONITOR WAS RESEATED TO THE MAIN CART CONNECTION WITH NO CHANGE. THE CASE WAS CONTINUED WITH THE FLICKERING MONITOR. THERE WAS NO REPORTED DELAY TO THE PROCEDURE DUE TO THIS ISSUE OR IMPACT ON PATIENT OUTCOME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6742481 ↗
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.