Medtronic, Inc. · BALLOON EUP22512X EUPHORA RX US · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169559950), received between Jun 16, 2017 and Jun 10, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (2 reports).

2 FDA recall records name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K143480); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 3, 2019Open, Classified

    Medtronic has determined that some catheters have a slightly higher diameter stylette causing removal difficulties.

    Z-0122-2020 · Root cause: Under Investigation by firm · FDA record ↗

  2. Jul 13, 2017Terminated

    Stylette removal difficulties on the Euphora and Solarice products.

    Z-3136-2017 · Root cause: Process design · FDA record ↗

Most reported problems

Device problems

  • Deflation Problem2
  • Difficult to Remove1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2024Deflation Problem; Difficult to Remove; Material Deformation; Deflation Problem; Difficult to Remove; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE EUPHORA CORONARY BALLOON CATHETER TO TREAT A MILDLY TORTUOUS, MILDLY CALCIFIED LESION WITH 100% STENOSIS IN THE MID CIRCUMFLEX (CX) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. THE BALLOON WAS BEING USED TO PRE-FILL. IT WAS REPORTED THAT BALLOON DEFLATION DIFFICULTIES OCCURRED, AND THE DEVICE WOULD NOT DEFLATE AT THE LESION SITE. REMOVAL DIFFICULTIES WERE ALSO REPORTED FOLLOWING INFLATION AS THE BALLOON WOULD NOT DEFLATE. THE DEVICE WAS ABLE TO BE PULLED OUT AS IT WAS SMALL. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19496131 ↗
  • MalfunctionJun 16, 2017Deflation Problem; Deflation Problem

    THE PHYSICIAN INTENDED TO USE A EUPHORA BALLOON CATHETER TO TREAT A LESION LOCATED IN THE LAD. NO DAMAGE WAS NOTED TO PACKAGING AND NO ISSUES WERE NOTED WHEN REMOVING THE DEVICES FROM THE TRAY/HOOP. THE PHYSICIAN NOTED THAT IT WAS DIFFICULT TO REMOVE THE STYLETTE. THE DEVICES WERE INSPECTED AND PREPPED WITH NO ISSUES IDENTIFIED. THE LESION WAS PRE-DILATED. IT WAS REPORTED THAT DEFLATION DIFFICULTIES WERE ENCOUNTERED. THE BALLOON WOULD NOT DEFLATE AT THE LESION SITE. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT, RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE AND NO EXCESSIVE FORCE WAS USED DURING DELIVERY. THE INDEFLATOR WAS TAKEN OFF THE BALLOON CATHETER AND THE BALLOON WAS DEFLATED WITH A SYRINGE. NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6646036 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K143480. One submission often covers many models, so check the product and lot details in each record.

  • Z-3137-2017 · Jul 13, 2017 · Terminated
    Solarice Rapid Exchange Balloon Dilatation Catheter, a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System.

Other Catheters, transluminal coronary angioplasty, percutaneous devices