Euphora™ EUP22520XRecalls & reported problems
Medtronic, Inc. · BALLOON EUP22520X EUPHORA RX US · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169559974), received between Jan 13, 2020 and Jan 20, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
2 FDA recall records name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K143480); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 3, 2019Open, Classified
Medtronic has determined that some catheters have a slightly higher diameter stylette causing removal difficulties.
Z-0122-2020 · Root cause: Under Investigation by firm · FDA record ↗
- Jul 13, 2017Terminated
Stylette removal difficulties on the Euphora and Solarice products.
Z-3136-2017 · Root cause: Process design · FDA record ↗
Most reported problems
Device problems
- Deflation Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 20, 2023Deflation Problem; Deflation Problem
AN ATTEMPT WAS MADE TO USE ONE EUPHORA RX PTCA BALLOON CATHETER TO TREAT A MILDLY TORTUOUS, MILDLY CALCIFIED LESION IN THE MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT BALLOON DEFLATION DIFFICULTIES OCCURRED. THE BALLOON WOULD NOT DEFLATE AT THE LESION SITE. THE PATIENT WAS REPORTED TO BE ALIVE WITH NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16213663 ↗MalfunctionJan 13, 2020Use of Device Problem; Use of Device Problem
DURING A PROCEDURE, AN ATTEMPT WAS MADE TO USE ONE EUPHORA PTCA BALLOON CATHETER TO TREAT A LESION. THERE WAS NO DAMAGE NOTED TO THE PACKAGING OR ANY ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP. IT WAS REPORTED THAT THE BALLOON WAS USED FEW DAYS AFTER THE EXPIRE DATE. THERE WAS NO PRODUCT ISSUE AND THE PATIENT WAS REPORTED TO BE ALIVE WITH NO INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9580748 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K143480. One submission often covers many models, so check the product and lot details in each record.
- Z-3137-2017 · Jul 13, 2017 · TerminatedSolarice Rapid Exchange Balloon Dilatation Catheter, a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System.
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page