Skip to content

Euphora™ EUP32512XRecalls & reported problems

Medtronic, Inc. · BALLOON EUP32512X EUPHORA RX US · Class 2

All models of Euphora (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K143480); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Sep 3, 2019Open, Classified

    Medtronic has determined that some catheters have a slightly higher diameter stylette causing removal difficulties.

    Z-0122-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jul 13, 2017Terminated

    Stylette removal difficulties on the Euphora and Solarice products.

    Z-3136-2017 · Root cause: Process design · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K143480. One submission often covers many models, so check the product and lot details in each record.

  • Z-3137-2017 · Jul 13, 2017 · Terminated
    Solarice Rapid Exchange Balloon Dilatation Catheter, a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System.

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page