Euphora™ EUP3515XRecalls & reported problems
Medtronic, Inc. · BALLOON EUP3515X EUPHORA RX US · Class 2
All models of Euphora (62)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169560284), received between Feb 16, 2023 and Mar 28, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
2 FDA recall records name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K143480); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Sep 3, 2019Open, Classified
Medtronic has determined that some catheters have a slightly higher diameter stylette causing removal difficulties.
Z-0122-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 13, 2017Terminated
Stylette removal difficulties on the Euphora and Solarice products.
Z-3136-2017 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Burst Container or Vessel2
- Difficult to Remove2
- Deflation Problem1
- Detachment of Device or Device Component1
- Inflation Problem1
- Material Deformation1
Patient problems
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 28, 2025Burst Container or Vessel; Deflation Problem; Inflation Problem; Difficult to Remove; Material Deformation; Burst Container or Vessel; Deflation Problem; Inflation Problem; Difficult to Remove; Material Deformation
IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING ONE EUPHORA RX CORONARY BALLOON CATHETER, BALLOON DEFLATION DIFFICULTIES OCCURRED. THE BALLOON WOULD NOT DEFLATE AT THE LESION SITE. IT WAS ALSO REPORTED THAT THERE WERE DIFFICULTIES REMOVING THE BALLOON FOLLOWING BALLOON INFLATION. THE DEFLATION DIFFICULTIES WERE NOTED FOLLOWING THE FIRST INFLATION AT 8 ATM. ATTEMPTS WERE MADE TO DEFLATE THE BALLOON BY PULLING NEGATIVE USING BOTH AN INSUFFLATOR AND A 20ML SYRINGE TO NO AVAIL. THE DEVICE WAS PULLED OUT OF THE VESSEL AND THROUGH THE BODY TO THE SHEATH IN THE RIGHT ARM WHERE THE BALLOON APPEARED TO BURST. THE DEVICE WAS THEN ABLE TO BE REMOVED THROUGH THE SHEATH. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21718028 ↗InjuryFeb 16, 2023Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component; Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component
AN ATTEMPT WAS MADE TO USE ONE EUPHORA RX PTCA BALLOON CATHETER TO TREAT A MODERATELY TORTUOUS, MILDLY CALCIFIED LESION IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. THE DEVICE WAS NOT KINKED AND RE-STRAIGHTENED DURING USE. IT WAS REPORTED THAT A BALLOON BURST OCCURRED DURING THE FIRST BALLOON INFLATION AT NOMINAL 8 ATM. IT WAS ALSO REPORTED THAT DURING REMOVAL OF THE DEVICE A BALLOON DETACHMENT OCCURRED. REMOVAL DIFFICULTIES WERE ENCOUNTERED WHEN ATTEMPTING TO REMOVE THE DETACHED PORTION FROM THE PATIENT. THE DETACHED PORTION OF THE DEVICE REMAINS IN THE PATIENT. THE PATIENT IS GOING FOR OPEN HEART SURGERY DUE TO THE BALLOON RUPTURE, TO REMOVE A PIECE OF THE BALLOON LEFT IN THE RCA. THE PATIENT IS ALIVE WITH NO FURTHER INJURY.
Patient problems: Device Embedded In Tissue or Plaque; Insufficient Information
MAUDE report 16387906 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K143480. One submission often covers many models, so check the product and lot details in each record.
- Z-3137-2017 · Jul 13, 2017 · TerminatedSolarice Rapid Exchange Balloon Dilatation Catheter, a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System.
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page