Skip to content

Euphora™ EUP3530XRecalls & reported problems

Medtronic, Inc. · BALLOON EUP3530X EUPHORA RX US · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169560314), received between May 13, 2020 and May 13, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

2 FDA recall records name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K143480); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Sep 3, 2019Open, Classified

    Medtronic has determined that some catheters have a slightly higher diameter stylette causing removal difficulties.

    Z-0122-2020 · Root cause: Under Investigation by firm · FDA record ↗

  2. Jul 13, 2017Terminated

    Stylette removal difficulties on the Euphora and Solarice products.

    Z-3136-2017 · Root cause: Process design · FDA record ↗

Most reported problems

Device problems

  • Deflation Problem1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2020Deflation Problem; Deflation Problem

    A EUPHORA RX BALLOON WAS INTENDED TO BE USED DURING A PROCEDURE TO TREAT A LESION IN THE MID RCA, EXHIBITING MODERATE TORTUOSITY AND MODERATE CALCIFICATION. STENOSIS WAS 90%. THERE WAS NO DAMAGE NOTED TO PACKAGING, AND THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE AND EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT IS REPORTED THAT BALLOON DEFLATION DIFFICULTIES OCCURRED - THE BALLOON WOULD NOT DEFLATE AT THE LESION SITE. THE PHYSICIAN REMOVED THE DEVICE WHILE INFLATED, AFTER MULTIPLE ATTEMPTS TO DEFLATE. NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10050823 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K143480. One submission often covers many models, so check the product and lot details in each record.

  • Z-3137-2017 · Jul 13, 2017 · Terminated
    Solarice Rapid Exchange Balloon Dilatation Catheter, a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System.

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.