Visia AF™ VR DVAB1D4 (UDI 00643169566354)Recalls & reported problems

Medtronic, Inc. · ICD-VR DVAB1D4 VISIA AF US DF4 · Class 3 · Implant

All models of Visia AF VR (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Visia AF implantable defibrillator detects heart rhythms and can deliver shocks. The 42 reports mostly describe sensing problems and inappropriate or missed shocks; some mention infections. They include 1 death and 33 injuries. There are 2 recalls: 1 is open because a subset may have less time between a replacement warning and the device reaching end of service, while the other is closed; reports fell from 6 in the previous 12 months to 1 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
1
Injuries reported
33
Recalls naming it
2

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 00643169566354), received between Aug 9, 2017 and Nov 10, 2025 — classified as 1 death, 33 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (16 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 3, 2021Class IOpen, Classified

    A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation

    Z-1305-2021 · Root cause: Device Design · Recall details · FDA record ↗

  2. Dec 19, 2017Class ITerminated

    Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).

    Z-0589-2018 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Sensing Problem16
  • Inappropriate/Inadequate Shock/Stimulation12
  • Adverse Event Without Identified Device or Use Problem10
  • Electromagnetic Interference3
  • Failure to Deliver Shock/Stimulation3
  • Pacing Problem3
  • Premature Discharge of Battery3
  • Signal Artifact/Noise3

Patient problems

  • Shock from Patient Lead(s)15
  • Unspecified Infection10
  • No Clinical Signs, Symptoms or Conditions8
  • Endocarditis3
  • No Known Impact Or Consequence To Patient3
  • Tachycardia2
  • Anxiety1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE HANDHELD DEVICE WAS NOT HOLDING A CHARGE. NO TROUBLESHOOTING INDICATED. THE MONITOR REMAINS IN USE. IT WAS ALSO NOTED BY THE PATIENT THAT THEY HAVE A ¿PROBLEM¿ WITH THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND IT WAS GOING OFF THE NIGHT PRIOR. ADDITIONALLY, IT WAS NOTED THAT THE PATIENT HAD RUNS OF VENTRICULAR TACHYCARDIA (VT) AND FELT LIGHTHEADED. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dizziness; Shock from Patient Lead(s); Tachycardia

    MAUDE report 23511086 ↗
  • InjurySep 2, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR) DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22945868 ↗
  • InjurySep 2, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22938824 ↗
  • InjuryAug 28, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR) DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22916149 ↗
  • InjuryAug 28, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22913820 ↗
  • InjuryJun 9, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE THERAPY FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT THE DEVICE DETECTED AS VENTRICULAR FIBRILLATION (VF). THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22169177 ↗
  • InjuryMay 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PATIENT EXPERIENCE AN INFECTION WITH VEGETATION ON THE RIGHT VENTRICULAR (RV) LEAD. THE DEVICE SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Medical device site infection; Unspecified Infection

    MAUDE report 22117127 ↗
  • InjurySep 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PATIENT EXPERIENCED AN SYSTEM INFECTION. THE PATIENT WAS TREATED WITH INTRAVENOUS (IV) ANTIBIOTIC MEDICATION  AND THE DEVICE SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 20257788 ↗
  • InjuryMay 8, 2024Signal Artifact/Noise; Mechanical Problem; Signal Artifact/Noise; Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A LEAD INTEGRITY ALERT (LIA) ON THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED BY NON-SUSTAINED VENTRICULAR TACHYCARDIA, AND SHORT V-V INTERVALS. NOISES WAS ALSO NOTED ON THE LEAD. IT WAS UNCLEAR IF THE NOISE WAS DUE TO A POTENTIAL LEAD FRACTURE OR DUE TO AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD HEADER ISSUE. THE ICD AND LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19268737 ↗
  • InjuryApr 12, 2024Failure to Deliver Shock/Stimulation; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT WAS IN VENTRICULAR ARRHYTHMIA AND PRESENTED TO THE EMERGENCY DEPARTMENT. THE PHYSICIAN WAS SPECULATING THAT THE PATIENT WAS NOT RECEIVING THERAPY FOR CURRENT VENTRICULAR TACHYCARDIA (VT). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Tachycardia

    MAUDE report 19097645 ↗
  • MalfunctionFeb 29, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT POST OPERATION THE DOCTOR THOUGHT THE  IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS NOT OPERATING PROPERLY DUE TO RATES IN THE 40'S BEATS PER MINUTE.  THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THISEVENT.

    Patient problems: Bradycardia

    MAUDE report 18809263 ↗
  • MalfunctionFeb 8, 2024Pacing Problem; Device Sensing Problem; Pacing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED CHEST PAIN AND THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE ANTI-TACHYCARDIA PACING FOR AN EPISODE OF SUPRA VENTRICULAR TACHYCARDIA (SVT) WHEN THE DEVICE MISCLASSIFIED THE EPISODE AS FAST VENTRICULAR TACHYCARDIA (FVT). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Chest Pain

    MAUDE report 18669865 ↗
  • InjuryJan 31, 2024Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED SHOCKS FOR DETECTED VENTRICULAR TACHYCARDIA (VT)/VENTRICULAR FIBRILLATION (VF). IT WAS SUSPECTED THAT THE ARRHYTHMIA MAY HAVE BEEN AN SUPRA VENTRICULAR TACHYCARDIA (SVT). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 18616309 ↗
  • InjuryJan 9, 2024Failure to Deliver Shock/Stimulation; Pacing Problem; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Pacing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED  INAPPROPRIATE THERAPY WITH ANTI-TACHYCARDIA PACING FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) VERSUS FAST VENTRICULAR TACHYCARDIA (VT). THERE WAS ALSO INAPPROPRIATE THERAPY WITHHELD  FOR SVT VERSUS VENTRICULAR TACHYCARDIA (VT). THE DEVICE REMAINS IN USE . NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 18479804 ↗
  • InjuryAug 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DISCOMFORT AND AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM INFECTION. THE DEVICE SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort; Unspecified Infection

    MAUDE report 17461323 ↗
  • InjuryDec 8, 2022Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THERE HAVE BEEN SEVERAL ALERTS ON THE RIGHT VENTRICULAR (RV) LEAD FOR LOW PACING IMPEDANCE. IT WAS ALSO REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED FAST VENTRICULAR TACHYCARDIA (FVT) AND DELIVERED MULTIPLE SHOCKS. IT WAS SUSPECTED THAT THE RHYTHM WAS ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR) AND IS THEREFORE INAPPROPRIATE. THE DEVICE AND LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 15942868 ↗
  • MalfunctionNov 23, 2022Battery Problem; Battery Problem

    THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO NORMAL BATTERY DEPLETION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15851251 ↗
  • MalfunctionOct 4, 2022Battery Problem; Battery Problem

    THE DEVICE WAS RETURNED TO THE MANUFACTURER WITH NO INFORMATION, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538211 ↗
  • InjurySep 30, 2022Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE DEVICE EXHIBITED PREMATURE BATTERY DEPLETION. IN JULY, THE EXPECTED BATTERY LONGEVITY WAS NEARLY SIX YEARS AND IN AUGUST THE EXPECTED LONGEVITY DECLINED TO NEARLY THREE YEARS REMAINING. THE DEVICE ALARMED RECENTLY FOR THE DEVICE REACHING RECOMMENDED REPLACEMENT TIME (RRT). THE PATIENT REPORTED BEING STRESSED, NERVOUS AND SCARED DUE TO THE DEVICE ISSUES. THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Anxiety

    MAUDE report 15520438 ↗
  • MalfunctionAug 18, 2022Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing; Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing

    IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT (LIA) WAS TRIGGERED DUE TO HIGH RATE NON-SUSTAINED EPISODES AND ELEVATED SENSING INTEGRITY COUNTER (SIC). IN ADDITION, AN EPISODE OF NOISE ON BOTH NEAR AND FAR FIELD ELECTROGRAMS (EGM) ALONG WITH OVERSENSING WAS OBSERVED. IT WAS CONFIRMED THAT THE PATIENT WAS UNDERGOING HIP SURGERY AT THE TIME OF THE STORED EPISODES. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15251347 ↗
  • InjuryJul 14, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS REPLACED NON-PROPHYLACTICALLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15014131 ↗
  • InjuryMar 29, 2022Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXPERIENCED AN UNEXPLAINED DROP IN BATTERY VOLTAGE AND TRIGGERED THE RECOMMENDED REPLACEMENT TIME (RRT) AND ELECTIVE REPLACEMENT INDICATOR (ERI) EARLIER THAT EXPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13938241 ↗
  • InjuryMar 25, 2022Failure to Deliver Shock/Stimulation; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED SUPRA VENTRICULAR TACHYCARDIA (SVT). IT WAS SUSPECTED THAT THE RHYTHM WAS FAST VENTRICULAR TACHYCARDIA (FVT) AND IS THEREFORE INAPPROPRIATELY WITHHELD THERAPY. THE DEVICE REMAINS IN USE AND APPEARS TO BE CURRENTLY FUNCTIONING AS PROGRAMMED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13902874 ↗
  • InjuryJan 14, 2022Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXPERIENCED RAPID BATTERY DEPLETION WHICH TRIGGERED THE RECOMMENDED REPLACEMENT TIME (RRT) EARLIER THAN EXPECTED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13258681 ↗
  • InjuryOct 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO AN (B)(6) INFECTION. THE PATIENT WAS INDICATED TO HAVE ENDOCARDITIS. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Endocarditis; Unspecified Infection

    MAUDE report 12704272 ↗

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →