Resolute Onyx™ RONYX22526WRecalls & reported problems

Medtronic, Inc. · STENT RONYX22526W RESOLUTE ONYX 2.25X26 · Class 3 · Implant

All models of Resolute Onyx (101)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169602267), received between Dec 13, 2017 and Mar 1, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration or Expulsion of Device2
  • Adverse Event Without Identified Device or Use Problem1
  • Leak/Splash1
  • Malposition of Device1

Patient problems

  • Myocardial Infarction3
  • Occlusion3
  • Thrombus3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 1, 2024Leak/Splash; Leak/Splash

    AN ATTEMPT WAS MADE TO USE ONE RESOLUTE ONYX OTW CORONARY DRUG-ELUTING STENT TO TREAT A MODERATELY TORTUOUS, SEVERELY CALCIFIED LESION IN THE DISTAL LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE DEVICE WAS INSPECTED, WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT LEAK OCCURRED AT THE LUER/HUB AT 12 ATMS. IT WAS DETAILED THAT TWO INFLATIONS WERE APPLIED TO THE DEVICE PRIOR TO THE ISSUE OCCURRING. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18815095 ↗
  • InjuryDec 13, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PHYSICIAN USED FOUR RESOLUTE ONYX OTW CORONARY DRUG ELUTING STENT TO TREAT A MILDLY TORTUOUS AND CALCIFIED LESION LOCATED IN THE PROXIMAL LEFT ANTERIOR DESCENDING ARTERY, EXHIBITING 100% STENOSIS. ARTERY DIAMETER 2.5 MM AND ARTERY LENGTH 24 MM. THE 2.5 X 18 STENT WAS DEPLOYED AT 18ATM FOR 8 SECONDS AND POST DILATED 3 TIMES IN THE DISTAL OF THE PROXIMAL LAD. THE 2.25 X 26 WAS DEPLOYED AT 18 ATM FOR 8 SECONDS AND POST DILATED 4 TIMES. THE 2.25 X 26 STENT WAS DEPLOYED AT 18 ATM FOR 10 SECONDS AND POST DILATED 5 TIMES, TWENTY (20) ATM FOR 8 SECONDS AND 6 TIMES 20 ATM FOR 6 SECONDS .THE 2.25 X 26 STENTS WERE CONFORMED TO BE DISTAL TO THE 2.5 X 18 HOWEVER, AS THE PATIENT HAD CHEST PAIN HE WAS MOVING AROUND, THE STENTS WERE SEEN TO BE PROXIMAL TO THE FIRST. DURING ANGIOGRAPHY, POST PROCEDURE, THE STENT HAD MOVED PROXIMALLY IN THE LAD. THE 2.25 X 15 WAS DEPLOYED AT 12ATM FOR 8 SECONDS AND POST DILATED 7 AT18 ATM FOR 6 SECONDS. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING, AND NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICES WERE INSPECTED WITH NO ISSUES IDENTIFIED. NEGATIVE PR

    Patient problems: Myocardial Infarction; Occlusion; Thrombus

    MAUDE report 7112115 ↗
  • InjuryDec 13, 2017Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT THE PHYSICIAN USED FOUR RESOLUTE ONYX OTW CORONARY DRUG ELUTING STENT TO TREAT A MILDLY TORTUOUS AND CALCIFIED LESION LOCATED IN THE PROXIMAL LEFT ANTERIOR DESCENDING ARTERY, EXHIBITING 100% STENOSIS. ARTERY DIAMETER 2.5 MM AND ARTERY LENGTH 24 MM. THE 2.5 X 18 STENT WAS DEPLOYED AT 18 ATM FOR 8 SECONDS AND POST DILATED 3 TIMES IN THE DISTAL OF THE PROXIMAL LAD. THE 2.25 X 26 WAS DEPLOYED AT 18 ATM FOR 8 SECONDS AND POST DILATED 4 TIMES. THE 2.25 X 26 STENT WAS DEPLOYED AT 18 ATM FOR 10 SECONDS AND POST DILATED 5 TIMES, 20 ATM FOR 8 SECONDS AND 6 TIMES 20 ATM FOR 6 SECONDS .THE 2.25 X 26 STENTS WERE CONFORMED TO BE DISTAL TO THE 2.5 X 18 HOWEVER, AS THE PATIENT HAD CHEST PAIN HE WAS MOVING AROUND, THE STENTS WERE SEEN TO BE PROXIMAL TO THE FIRST. DURING ANGIOGRAPHY, POST PROCEDURE, THE STENT HAD MOVED PROXIMALLY IN THE LAD. THE 2.25 X 15 WAS DEPLOYED AT 12 ATM FOR 8 SECONDS AND POST DILATED 7 AT 18 ATM FOR 6 SECONDS. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING, AND NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICES WERE INSPECTED WITH NO ISSUES IDENTIFIED. NEGATIVE PREP WAS

    Patient problems: Myocardial Infarction; Occlusion; Thrombus

    MAUDE report 7112109 ↗
  • InjuryDec 13, 2017Migration or Expulsion of Device; Malposition of Device; Migration or Expulsion of Device; Malposition of Device

    IT WAS REPORTED THAT THE PHYSICIAN USED FOUR RESOLUTE ONYX OTW CORONARY DRUG ELUTING STENT TO TREAT A MILDLY TORTUOUS AND CALCIFIED LESION LOCATED IN THE PROXIMAL LEFT ANTERIOR DESCENDING ARTERY, EXHIBITING 100% STENOSIS. ARTERY DIAMETER 2.5 MM AND ARTERY LENGTH 24 MM. THE 2.5 X 18 STENT WAS DEPLOYED AT 18 ATM FOR 8 SECONDS AND POST DILATED 3 TIMES IN THE DISTAL OF THE PROXIMAL LAD. THE 2.25 X 26 WAS DEPLOYED AT 18 ATM FOR 8 SECONDS AND POST DILATED 4 TIMES. THE 2.25 X 26 STENT WAS DEPLOYED AT 18 ATM FOR 10 SECONDS AND POST DILATED 5 TIMES, 20 ATM FOR 8 SECONDS AND 6 TIMES 20 ATM FOR 6 SECONDS .THE 2.25 X 26 STENTS WERE CONFORMED TO BE DISTAL TO THE 2.5 X 18 HOWEVER, AS THE PATIENT HAD CHEST PAIN HE WAS MOVING AROUND, THE STENTS WERE SEEN TO BE PROXIMAL TO THE FIRST. DURING ANGIOGRAPHY, POST PROCEDURE, THE STENT HAD MOVED PROXIMALLY IN THE LAD. THE 2.25 X 15 WAS DEPLOYED AT 12 ATM FOR 8 SECONDS AND POST DILATED 7 AT 18 ATM FOR 6 SECONDS. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING, AND NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICES WERE INSPECTED WITH NO ISSUES IDENTIFIED. NEGATIVE PREP WAS

    Patient problems: Myocardial Infarction; Occlusion; Thrombus

    MAUDE report 7112082 ↗