Skip to content

Resolute Onyx™ RONYX25012WRecalls & reported problems

Medtronic, Inc. · STENT RONYX25012W RESOLUTE ONYX 2.50X12 · Class 3 · Implant

All models of Resolute Onyx (101)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169602311), received between Jan 3, 2019 and Feb 13, 2024 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Leak/Splash2
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Myocardial Infarction1
  • No Consequences Or Impact To Patient1
  • Occlusion1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2024Leak/Splash; Leak/Splash

    AN ATTEMPT WAS MADE TO USE ONE RESOLUTE ONYX OTW CORONARY DRUG ELUTING STENT TO TREAT A MODERATELY CALCIFIED, MODERATELY TORTUOUS LESION WITH 80% STENOSIS IN THE MID RIGHT CORONARY ARTERY (RCA). THE DEVICE WAS NOT INSPECTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT THERE WAS A SLOW LEAK. THE STENT WAS IMPLANTED. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18693324 ↗
  • MalfunctionSep 19, 2022Leak/Splash; Leak/Splash

    AN ATTEMPT WAS MADE TO USE ONE RESOLUTE ONYX OTW CORONARY DRUG ELUTING STENT (DES) TO TREAT A MODERATELY TORTUOUS, MODERATELY CALCIFIED LESION LOCATED IN THE MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE DEVICE WAS INSPECTED PRIOR TO USE WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT WHEN THE BALLOON WAS INFLATED THERE WAS A SLOW DRIP OR LEAK AT THE TRANSITION OF THE CATHETER AND THE HUB. THE PATIENT IS REPORTED TO BE ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15445199 ↗
  • InjuryJan 6, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING THE INDEX PROCEDURE A RESOLUTE ONYX DRUG ELUTING STENT WAS IMPLANTED IN THE RCA. IT WAS REPORTED THAT THE PATIENT SUFFERED A SPASM ON THE SAME DAY AS THE INDEX PROCEDURE. IT WAS REPORTED THAT THE VASOSPASM WAS NOT TREATED BUT CONSIDERED SIGNIFICANT BY INVESTIGATOR.

    Patient problems: Occlusion

    MAUDE report 9552497 ↗
  • MalfunctionMar 28, 2019Inflation Problem; Inflation Problem

    DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE A RESOLUTE ONYX RX CORONARY DRUG ELUTING STENT TO TREAT A MILDLY TORTUOUS LESION LOCATED IN THE MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE AND EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT THERE WERE INFLATION DIFFICULTIES EXPERIENCED DURING STENT DEPLOYMENT WHILE INFLATING AT 12ATM. ANOTHER INFLATION DEVICE WAS USED TO ATTEMPT INFLATING THE DEVICE HOWEVER IT WOULD STILL NOT INFLATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER 2.50X12MM RESOLUTE ONYX STENT. THE PATIENT WAS REPORTED TO BE ALIVE WITH NO INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8459437 ↗
  • InjuryJan 3, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ONE RESOLUTE ONYX DRUG ELUTING STENT WAS IMPLANTED IN THE LAD IN A PATIENT. ONE DAY POST INDEX, ELEVATED ENZYMES WERE NOTED WHICH IS REPORTED TO BE RELATED TO MI, THE MI WAS IN A NON TARGET VESSEL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO DEVICE OR TO THE ANTI PLATELET. PATIENT RECOVERED.

    Patient problems: Myocardial Infarction

    MAUDE report 8211802 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page