Skip to content

Resolute Onyx™ RONYX25026WRecalls & reported problems

Medtronic, Inc. · STENT RONYX25026W RESOLUTE ONYX 2.50X26 · Class 3 · Implant

All models of Resolute Onyx (101)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169602359), received between May 24, 2021 and May 24, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Unspecified Tissue Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 24, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE ONE RESOLUTE ONYX RX CORONARY DRUG ELUTING STENT TO TREAT A SEVERELY CALCIFIED LESION IN THE PROXIMAL/ OSTIUM LM/LAD ARTERY WITH 90-95% STENOSIS. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS NOT PERFORMED. THE LESION WAS PREDILATED WITH A 2.5MM X 15 BALLOON. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED DURING DELIVERY TO/AT LESION. A SMALL BALLOON WAS INSERTED TO RETRIEVE DISLODGED STENT HOWEVER WAS UNSUCCESSFUL. A SNARE WAS THEN USED TO RETRIEVE STENT WITHOUT SUCCESS. THE PATIENT WAS HEMODYNAMICALLY STABLE WHEN TRANSFERRED TO ANOTHER FACILITY TO RETRIEVE THE STENT AND FINISH THE PROCEDURE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11869539 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page