TYRX™ Neuro Absorbable Antibacterial Envelope - Large NMRM6133Recalls & reported problems
Medtronic, Inc. · ENVELOPE NMRM6133 NEURO ABSORB LRG US · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169624795), received between Dec 9, 2016 and Oct 3, 2018 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Device Contamination with Chemical or Other Material1
- Material Integrity Problem1
Patient problems
- Discomfort1
- No Known Impact Or Consequence To Patient1
- Tissue Damage1
- Unspecified Infection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 3, 2018Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT THAT HAD A DEEP BRAIN STIMULATION IMPLANTABLE PULSE GENERATOR (IPG) IMPLANTED ALONG WITH A ABSORBABLE ANTIBACTERIAL ENVELOPE. THE PATIENT HAD REPORTED DISCOMFORT AT THE AT THE IMPLANT SITE AND APPROXIMATELY NINE MONTHS POST IMPLANT A POCKET REVISION WAS COMPLETED. WHEN THE SURGEON OPENED THE INCISION THEY FOUND THAT THE ABSORBABLE ANTIBACTERIAL ENVELOPE HAD FUSED TO THE UNDERLYING TISSUE. THE SURGEON HAD TO CUT AWAY THE TISSUE AND EXTENSION IN ORDER TO REMOVE THE ENVELOPE. THE IPG WAS PLACED BACK INTO THE POCKET AND AN ANTIBIOTIC POWDER WAS PLACED IN THE POCKET AND THE INCISION WAS CLOSED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Discomfort; Tissue Damage
MAUDE report 7930662 ↗InjuryJun 8, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT APPROXIMATELY THREE WEEKS AFTER IMPLANT THE PATIENT DEVELOPED A POCKET INFECTION. THE ABSORBABLE ANTIBACTERIAL ENVELOPE WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 6624725 ↗MalfunctionDec 9, 2016Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE ABSORBABLE ENVELOPE WAS OPENED AND FOUND TO HAVE A VISIBLE ANOMALY ON THE ENVELOPE, A BLACK SPEC. THE ENVELOPE WAS INSPECTED BEFORE GOING TO THE STERILE FIELD BY THE IMPLANTING PHYSICIAN WHO DETERMINED IT SHOULD BE DISCARDED. ANOTHER ENVELOPE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6164742 ↗
Other Mesh, surgical, polymeric devices
- PROCEED PVPMJohnson & Johnson International Inc. · Mesh, surgical, polymeric1,233 reports
- PROCEED PCDN1Johnson & Johnson International Inc. · Mesh, surgical, polymeric744 reports
- PROCEED PCDG1Johnson & Johnson International Inc. · Mesh, surgical, polymeric1 recall1 death637 reports
- PROCEED PVPSJohnson & Johnson International Inc. · Mesh, surgical, polymeric622 reports
- ProGrip LPG1510Sofradim Production · Mesh, surgical, polymeric2 deaths616 reports
- VENTRALEX 0010302Davol Inc. · Mesh, surgical, polymeric4 deaths531 reports
- BARD 0112680Davol Inc. · Mesh, surgical, polymeric1 death494 reports
- ProGrip TEM1509GSofradim Production · Mesh, surgical, polymeric3 deaths471 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page