Skip to content

T2 STRATOSPHERE Expandable Corpectomy System 436113DRecalls & reported problems

Medtronic Sofamor Danek, Inc. · CENTERPIECE 436113D 13MM D-SIZE STERILE · Class 2 · Implant

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169634824), received between Feb 14, 2022 and Dec 7, 2023 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Mechanics Altered” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanics Altered2
  • Activation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Loss of consciousness1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 7, 2023Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH SPINAL THERAPY. IT WAS REPORTED THAT COULDN'T LOCK CAGE BECAUSE SET SCREW WAS BINDED UP. STERILE PACKED IMPLANT. SURGERY DELAYED BY 45 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18280486 ↗
  • MalfunctionJul 29, 2022Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A CAGE AND END CAP TO BE IMPLANTED IN A PATIENT DURING A VERTEBRAL BODY REPLACEMENT FOR AN L1 VERTEBRAL BODY COMPRESSION. IT WAS REPORTED THAT F13D 16 MM CAP WAS INSERTED IN L1 AND INSTALLED IN THE VERTEBRAL BODY. IT WAS ABLE TO BE INSTALLED, EXPANDED, AND THE FINAL TIGHTENING WAS PERFORMED. T2 ITSELF ALSO MOVED A LITTLE WHEN REMOVING THE CAGE FROM THE HOLDER, SO IT WAS REMOVED. THE REMOVED CAGE WAS PERFORMING A FINAL, BUT IT WAS EXPANDED WHEN THE PHYSICIAN HELD IT. SINCE IT WAS EXPANDED DESPITE BEING FINAL, IT WAS NOT REINSTALLED AND WAS REPLACED WITH F20 CAGE. THE CENTERPIECE COULD NOT BE FASTENED IN THE PATIENT'S BODY. THE END CAP WAS TOO HARD TO BE ATTACHED. THE PRODUCTS WERE USED ON THE PATIENT AND THE PATIENT WAS ON THE OPERATING TABLE WHEN THE PRODUCTS MALFUNCTIONED. THERE WERE NO SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT. THERE WAS A DELAY OF LESS THAN 60 MINUTES IN OVERALL PROCEDURE TIME. IT WAS UNKNOWN IF THE PATIENT NEEDED TO BE HOSPITALIZED/ STAY LONGER. THE DATES OF IMPLANT AND EXPLANT WAS (B)(6) 2022. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15127510 ↗
  • InjuryFeb 14, 2022Activation Problem; Activation Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH A CENTERPIECE AND ANTERIOR PLATE ZEVO IN A SOMATECTOMY. IT WAS REPORTED THAT DURING THE FIRST SURGERY ON (B)(6) 2022, THE CAGE DIDN'T MAINTAIN THE EXPANSION, SO IT WAS REPLACED WITH ANOTHER CAGE USING THE TWO ENDPLATES TO GIVE LORDOSIS. THE ENDPLATE HAD DIFFICULTIES TO ASSEMBLE THE CENTRAL PIECE DURING THE SAME SURGERY. 1 DAY POST-OP, THE PATIENT REPORTED THE SENSATION OF ¿CRUNCHING¿ AT THE CERVICAL LEVEL. FOLLOWING DIAGNOSTIC IMAGING, IT WAS FOUND THAT THE CRANIAL ENDPLATE HAD LOST THE LORDOSIS DETERMINED DURING THE SURGERY. THE DYNAMOMETRIC CLOSURE OF THE NUT INHERENT IN THE DEFINITIVE POSITIONING OF THE CRANIAL ENDPLATE TOOK PLACE WITHOUT ANY PROBLEM. NO REVISION WAS MADE. ON THE (B)(6) 2022, IT WAS REPORTED THAT FOLLOWING A DOMESTIC QUARREL WITHOUT A COLLAR, THE PATIENT UNDERWENT TWO FRONTAL FALLS WITH RELATIVE LOSS OF CONSCIOUSNESS. AFTER RECOVERING FROM THE SECOND FALL, TETRAPLEGIA WAS FOUND. UPON ARRIVAL IN THE FACILITY, THE DIAGNOSTIC IMAGES REVEALED IMPLANT MOBILIZATION. ON (B)(6) 2022, A REVISION SURGERY WAS MADE TO REMOVE THE MEDICAL DEVICES. WHI

    Patient problems: Loss of consciousness

    MAUDE report 13521310 ↗

Other Spinal vertebral body replacement device devices

All Spinal vertebral body replacement device devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.