T2 STRATOSPHERE Expandable Corpectomy System 436120BRecalls & reported problems
Medtronic Sofamor Danek, Inc. · CENTERPIECE 436120B 20MM B-SIZE STERILE · Class 2 · Implant
All models of T2 STRATOSPHERE Expandable Corpectomy System (52)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169634855), received between Dec 29, 2019 and Jan 5, 2022 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Migration2
- Break1
- Migration or Expulsion of Device1
Patient problems
- No Known Impact Or Consequence To Patient3
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA FIELD REPRESENTATIVE REGARDING PATIENT WITH UNKNOWN INDICATION INVOLVED IN THE REPORTED EVENT. LEVELS IMPLANTED: T10-L5. IT WAS REPORTED THAT THEY PERFORMED VERTEBRAL BODY REPLACEMENT AND POSTERIOR FUSION OF T11.12.L2 FOR L1 BURST. AFTER THAT, THE T3 VERTEBRAL BODY REPLACEMENT CAGE SUNK TO THE L2 VERTEBRAL BODY, AND AT THE SAME TIME, THE L2 SCREW BACKED OUT. THE L2 SCREW WAS REMOVED AND SINCE THERE WAS BKP IN L3, SCREW WAS ADDED TO L4.5 AND T10 TO EXTEND AND RE-FIXED. THERE WAS NO MALFUNCTION WITH THE PRODUCTS. PATIENT SYMPTOMS- DORSAL PAIN WAS REPORTED. PATIENT WAS HOSPITALIZED PROLONG AS A RESULT OF THIS EVENT.
Patient problems: Pain
MAUDE report 13168723 ↗InjuryJul 1, 2020Migration; Migration
PATIENT DEMOGRAPHICS: GENDER- FEMALE PRE OP DIAGNOSIS: VERTEBRAL BODY BURST FRACTURE PROCEDURE PERFORMED: L2 REPLACEMENT OF VERTEBRAL BODY, T12-L4 POSTERIOR PPS LEVELS IMPLANTED: L2 POSTERIOR T12-L4 IMPLANT DATE: (B)(6) 2020; EXPLANT DATE: (B)(6) 2020 (PLANNED) DEVICE STATUS: EXPLANT PLANNED IT WAS REPORTED THAT POST OP, CAGE BACKED OUT. THE CAGE, WHICH WAS INSERTED ABOUT 2 WEEKS, BACKED OUT AFTER THE OPERATION. THE EXTENT OF BACKING OUT WAS UNKNOWN. IT WAS PLANNED TO PERFORM A REOPERATION ON (B)(6) 2020 (PLAN TO REMOVE THE CAGE THAT HAS BACKED OUT AND REPLACE IT WITH T2). THE POSTERIOR SIDE HAS BEEN FIXED WITH V4, AND THE ANTERIOR VERTEBRAL BODY REPLACEMENT WAS PERFORMED AFTER THE POSTERIOR FIXATION. THE CAGE ON THE CRANIAL SIDE WAS PLACED IN THE DIRECTION OF THE ADJUSTING END CAP. PATIENT COMPLICATIONS WERE UNKNOWN. THE CAGE BACKED OUT AFTER THE OPERATION, SO HOSPITALIZATION WAS PROLONGED FOR THE REVISION SURGERY. IN REVISION SURGERY PERFORMED T2 STRATOSPHERE WAS REMOVED AND WAS REPLACED WITH T2 ALTITUDE. THERE WAS PATIENT COMPLICATION OCCURRED WHICH NEED TO PERFORM A RE-OPERATION. IT SEEMED THAT THE CAGE HAD BEEN PLACED WITH THE UPPER LEVEL DISC REMAINING. UPDATE RECEIVED ON THR
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10216971 ↗InjuryMay 7, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH L4 OSTEOPOROTIC VERTEBRAL FRACTURE; AND UNDERWENT ANTERIOR REPLACEMENT OF L4 VERTEBRAL BODY. THE ANGLE OF THE IMPLANT WAS FIXED AT AN APPROPRIATE ANGLE. ON AN UNKNOWN DATE, POST-OP, THE ANGLE OF THE ADJUSTABLE END CAP APPEARED TO HAVE CHANGED. THE IMPLANT WAS CONSIDERED TO HAVE MIGRATED. HENCE, A REVISION SURGERY WAS PERFORMED, IN WHICH THE IMPLANT WAS REMOVED AND WAS REPLACED WITH A NEW ONE (MANUFACTURED BY ANOTHER COMPANY). THE SURGEON REPORTED THAT THE PATIENT IS RECOVERING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10033027 ↗MalfunctionDec 29, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH PSEUDOARTHROSIS AFTER OSTEOPOROTIC VERTEBRAL FRACTURE AT T12. SHE UNDERWENT SUBTOTAL REMOVAL AT T12; AND ANTERIOR FIXATION AT T11-L1. INTRA-OP, WHEN ATTEMPTING TO PERFORM FINAL TIGHTENING FOR THE SET SCREW ON THE CRANIAL SIDE, THE SET SCREW COULD NOT BE CONFIRMED UNDER DIRECT VISION. THEN, THE SALES REP HAD A CONVERSATION WITH THE SURGEON, AND IT WAS DECIDED TO REMOVE THE INSERTER FIRST AND THEN TO PERFORM FINAL TIGHTENING FOR THE SET SCREW ON THE CRANIAL SIDE. SO, WHEN TRYING TO REMOVE THE INSERTER FROM THE CAGE, THE CAGE WAS ALSO REMOVED OUT OF THE OPERATION FIELD. WHEN TRYING TO LOOSEN THE SET SCREW IN THE CENTER, IT DID NOT COME LOOSE. HENCE, A REMOVAL TOOL WAS USED, AND THE EXTEND CAP WAS REMOVED. AT THAT TIME, THE CAGE BROKE INTO TWO PARTS WITH A CRISP SOUND. THE BROKEN CAGE WAS COMPLETELY OUT OF THE PATIENT WITH NO FRAGMENTS OF THE CAGE REMAINING INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NEW CAGE. THERE WAS A DELAY OF LESS THAN 60 MINUTES IN OVERALL PROCEDURE TIME AS A RESULT OF THIS EVENT BUT NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9528632 ↗
FDA clearance and classification
Cleared through Special 510(k) K183510 on Jan 16, 2019 (decision: Substantially Equivalent). Submitted by Medtronic Sofamor Danek USA, Inc. as “T2 STRATOSPHERE Expandable Corpectomy System”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code MQP (spinal vertebral body replacement device) under 21 CFR 888.3060 as a Class II device. 21 CFR 888.3060 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the T2 STRATOSPHERE Expandable Corpectomy System 436120B
Has the T2 STRATOSPHERE Expandable Corpectomy System 436120B been recalled?
No FDA recall record in our data names this exact device identifier (DI 00643169634855), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the T2 STRATOSPHERE Expandable Corpectomy System 436120B?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Dec 29, 2019 and Jan 5, 2022. In those reports, the most common specific device problems are “Migration” (2), “Break” (1) and “Migration or Expulsion of Device” (1); the most common patient problem is “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the T2 STRATOSPHERE Expandable Corpectomy System 436120B?
None of the reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the T2 STRATOSPHERE Expandable Corpectomy System 436120B FDA approved?
It was cleared through Special 510(k) K183510 on Jan 16, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the T2 STRATOSPHERE Expandable Corpectomy System 436120B MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the T2 STRATOSPHERE Expandable Corpectomy System 436120B, and is it still sold?
The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand T2 STRATOSPHERE Expandable Corpectomy System, model 436120B). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Spinal vertebral body replacement device devices
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7220018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 33-54 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 5-8 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7220016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page