T2 STRATOSPHERE Expandable Corpectomy System 436120450Recalls & reported problems
Medtronic Sofamor Danek, Inc. · ENDCAP 436120450 EXTENDED 20 X 45MM STER · Class 2 · Implant
All models of T2 STRATOSPHERE Expandable Corpectomy System (52)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00643169638426), received between Dec 25, 2019 and Jul 15, 2020 — classified as 6 injuries.
The most frequently reported specific device problem is “Migration” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration3
- Migration or Expulsion of Device2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Clinical Signs, Symptoms or Conditions2
- Bone Fracture(s)1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryJul 15, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT L3 VERTEBRAL BODY REPLACEMENT SURGERY DUE TO VERTEBRAL FRACTURE. THE PATIENT WAS SCHEDULED TO HAVE AN OPERATION THAT SET T3 AT L3 AND PERFORM OLIF AT L4/5 ON (B)(6) AND PERFORM PLIF AT L5/S AND POSTERIOR FIXATION ON (B)(6). DURING HOSPITALIZATION AFTER (B)(6) 2020 OPERATION, THE PATIENT WORE A CORSET AND WALKED TO A SEATED MEAL AND THE TOILET. POSTOPERATIVE PHOTOGRAPH WAS TAKEN ON (B)(6) 2020 AND IT SEEMED THAT THE IMPLANTED END CAP WAS ALREADY TILTED COMPARED TO THE STATE DURING SURGERY. ON (B)(6) WHILE TAKING RADIOGRAPHIC IMAGE DURING THE PREOPERATIVE DIAGNOSIS, IT WAS FOUND THAT THE IMPLANTED CAGE HAD BACKED OUT/MIGRATED. THE ANGLE OF THE END CAP HAD CHANGED. IN THE OPERATION PERFORMED ON (B)(6) 2020, IT WAS PLANNED TO PERFORM PLIF AT L5/S AND POSTERIOR FIXATION, BUT IT WAS NECESSARY TO REPLACE THE T3 CAGE, SO THE CAGE WAS REPLACED. IT WAS POSSIBLE THAT THE CONTRALATERAL DISC CURETTAGE WAS NOT PERFORMED PERFECTLY, BUT AS PER THE SURGEON OPINION, THE VERTEBRAL BODY WAS RAISED WELL DURING THE INTRA-OPERATIVE LENGTHENING.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10274627 ↗InjuryJul 15, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT L3 VERTEBRAL BODY REPLACEMENT SURGERY DUE TO VERTEBRAL FRACTURE. THE PATIENT WAS SCHEDULED TO HAVE AN OPERATION THAT SET T3 AT L3 AND PERFORM OLIF AT L4/5 ON (B)(6) AND PERFORM PLIF AT L5/S AND POSTERIOR FIXATION ON (B)(6). DURING HOSPITALIZATION AFTER (B)(6) 2020 OPERATION, THE PATIENT WORE A CORSET AND WALKED TO A SEATED MEAL AND THE TOILET. POSTOPERATIVE PHOTOGRAPH WAS TAKEN ON (B)(6) 2020 AND IT SEEMED THAT THE IMPLANTED END CAP WAS ALREADY TILTED COMPARED TO THE STATE DURING SURGERY. ON (B)(6), WHILE TAKING RADIOGRAPHIC IMAGE DURING THE PREOPERATIVE DIAGNOSIS, IT WAS FOUND THAT THE IMPLANTED CAGE HAD BACKED OUT/MIGRATED. THE ANGLE OF THE END CAP HAD CHANGED. IN THE OPERATION PERFORMED ON (B)(6) 2020, IT WAS PLANNED TO PERFORM PLIF AT L5/S AND POSTERIOR FIXATION, BUT IT WAS NECESSARY TO REPLACE THE T3 CAGE, SO THE CAGE WAS REPLACED. IT WAS POSSIBLE THAT THE CONTRALATERAL DISC CURETTAGE WAS NOT PERFORMED PERFECTLY, BUT AS PER THE SURGEON OPINION, THE VERTEBRAL BODY WAS RAISED WELL DURING THE INTRA-OPERATIVE LENGTHENING.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10274621 ↗InjuryJun 16, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH L4 OSTEOPOROTIC VERTEBRAL FRACTURE; AND UNDERWENT ANTERIOR REPLACEMENT OF L4 VERTEBRAL BODY AND ANTERIOR LUMBAR FIXATION. THE ANGLE OF THE END CAP WAS FIXED AT AN APPROPRIATE ANGLE. ON AN UNKNOWN DATE, POST-OP, THE ANGLE OF THE ADJUSTABLE END CAP APPEARED TO HAVE CHANGED. THE IMPLANT WAS CONSIDERED TO HAVE MIGRATED. HENCE, A REVISION SURGERY WAS PERFORMED, IN WHICH THE IMPLANT WAS REMOVED AND WAS REPLACED WITH A NEW ONE (MANUFACTURED BY ANOTHER COMPANY). THE SURGEON REPORTED THAT THE PATIENT IS RECOVERING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10160810 ↗InjuryMar 25, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT VERTEBRAL BODY ANTERIOR REPLACEMENT AT L2 DUE TO COMPRESSION FRACTURE OF VERTEBRAL BODY. POST-OP, IT WAS FOUND THAT THE CAGE HAS BACKED OUT BY CHECKING X-RAY IMAGE AFTER 2 WEEKS OF OPERATION. THE PATIENT WAS ABLE TO WALK WITH A CANE AND THERE WAS NO HEALTH DAMAGE IN THE PATIENT AT THIS POINT. HOWEVER, IT WAS CHECKED BY IMAGE THAT THE EXTENDED ENDCAP HAS CONTACTED WITH THE KIDNEY AND IT WAS FORECASTED THAT SERIOUS DAMAGE WILL OCCUR IF THE CAGE CONTACTS WITH THE ANTERIOR MAJOR BLOOD VESSEL. BONE QUALITY WAS VERY BAD. IT WAS FOUND BY CHECKING THE X-RAY IMAGE TAKEN IN THE FIRST WEEK AFTER THE OPERATION THAT THE VERTEBRAL BODY OF THE MOST UPPER LEVEL FRACTURED. MAYBE THE SCREW AT L4 LOOSENED. HENCE AN REVISION SURGERY PERFORMED ON (B)(6) 2020 IN WHICH ALL OF THE SCREWS ON THE POSTERIOR SIDE WERE REMOVED AND TO REPLACE THE SCREWS WITH THICK SCREWS AND AFTER THAT, RETRIEVE THE CAGE WHICH BACKED OUT TO THE ANTERIOR SIDE AND TO INSERT AUTOLOGOUS BONE.
Patient problems: Bone Fracture(s)
MAUDE report 9881765 ↗InjuryDec 25, 2019Migration or Expulsion of Device; Migration or Expulsion of Device
THE ADJUSTABLE END CAP HAD NOT BEEN LOCKED AFTER FINAL TIGHTENING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9517755 ↗InjuryDec 25, 2019Migration or Expulsion of Device; Migration or Expulsion of Device
THE ADJUSTABLE END CAP HAD NOT BEEN LOCKED AFTER FINAL TIGHTENING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9517752 ↗
Other Spinal vertebral body replacement device devices
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7160012Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7220018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7160014Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page