Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169650992), received between Sep 16, 2022 and Nov 19, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Peripheral Nervous Injury2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 19, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING LUMBAR 2 VERTEB RAL BODY RESECTION & SCREW-ROD INTERNAL FIXATION & PROSTHESIS REPLACEMENT FOR AN INDICATION OF LUMBAR METASTATIC TUMOR. IT WAS REPORTED THAT THE END CAP WAS BROKEN DURING THE OPERATION. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23592009 ↗InjurySep 11, 2025Migration; Migration
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED POST A CORPECTOMY PROCEDURE IN A PATIENT DIAGNOSED WITH TUMOR. IT WAS REPORTED THAT THE PROSTHETIC VERTEBRAL BODY LOOSENED AND MIGRATED. THE CAGE AND ENDCAP WERE EXPLANTED. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23028784 ↗InjurySep 16, 2022Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ANTEROLATERAL C ORPECTOMY FOR BURST FRACTURE. LEVELS IMPLANTED: L4 IT WAS REPORTED THAT THAT DEVICES BROKE AND WERE EXPLANTED. NO FRAGMENTS WERE LEFT IN THE PATIENT. THERE WAS NO PATIENT SYMPTOM REPORTED. IMMEDIATELY POST OP, THE PATIENT WAS NEUROLOGICALLY INTACT. HOWEVER, AFTER UNDERGOING PHYSICAL THERAPY, THE PATIENT DEVELOPED SOME NEUROLOGICAL DEFICIT IN THE LEFT LOWER EXTREMITY. THE PATIENT WAS BROUGHT BACK IN FOR IMAGING AND THE INSTRUMENTATION WAS INTACT. IT WAS THE SURGEONS BELIEF THAT THE PATIENT HAS DEVELOPED A PALSY DUE TO THE EXTENDED RETRACTION ON THE PSOAS MUSCLE. ONE OF THE CONTRIBUTING FACTORS IN THE AMOUNT OF TIME THE PSOAS WAS UNDER RETRACTION WAS DUE TO THE ISSUES WITH THE EXPANDABLE CORPECTOMY DEVICE. THESE ISSUES INCREASED THE AMOUNT OF TIME IT TOOK TO COMPLETE THE CORPECTOMY BY ABOUT AN HOUR. ALL THREE DEVICES CONTRIBUTED TO THE DELAY: THE EXTRA TIME TO REMOVE AND THEN ATTEMPT TO RESNAP THEM BACK ON, ETC. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: Peripheral Nervous Injury
MAUDE report 15436865 ↗InjurySep 16, 2022Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ANTEROLATERAL C ORPECTOMY FOR BURST FRACTURE. LEVELS IMPLANTED: L4 IT WAS REPORTED THAT THAT DEVICES BROKE AND WERE EXPLANTED. NO FRAGMENTS WERE LEFT IN THE PATIENT. THERE WAS NO PATIENT SYMPTOM REPORTED. IMMEDIATELY POST OP, THE PATIENT WAS NEUROLOGICALLY INTACT. HOWEVER, AFTER UNDERGOING PHYSICAL THERAPY, THE PATIENT DEVELOPED SOME NEUROLOGICAL DEFICIT IN THE LEFT LOWER EXTREMITY. THE PATIENT WAS BROUGHT BACK IN FOR IMAGING AND THE INSTRUMENTATION WAS INTACT. IT WAS THE SURGEONS BELIEF THAT THE PATIENT HAS DEVELOPED A PALSY DUE TO THE EXTENDED RETRACTION ON THE PSOAS MUSCLE. ONE OF THE CONTRIBUTING FACTORS IN THE AMOUNT OF TIME THE PSOAS WAS UNDER RETRACTION WAS DUE TO THE ISSUES WITH THE EXPANDABLE CORPECTOMY DEVICE. THESE ISSUES INCREASED THE AMOUNT OF TIME IT TOOK TO COMPLETE THE CORPECTOMY BY ABOUT AN HOUR. ALL THREE DEVICES CONTRIBUTED TO THE DELAY: THE EXTRA TIME TO REMOVE AND THEN ATTEMPT TO RESNAP THEM BACK ON, ETC. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: Peripheral Nervous Injury
MAUDE report 15436851 ↗
Other Spinal vertebral body replacement device devices
- X-CORE 7180018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7220018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 5-8 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7160012Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7180016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7160014Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page