T2 STRATOSPHERE™ Expandable Corpectomy System 436013D (UDI 00643169663145)Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CENTERPIECE 436013D 13MM D-SIZE · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169663145), received between May 25, 2021 and Jul 25, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Activation Failure1
- Material Deformation1
- Mechanical Jam1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2025Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER, DISTRIBUTOR) REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT DURING THE PLACEMENT OF THE CAGE, THE SCREW DRIVER SPUN FREELY AND WOULD NOT TIGHTEN WHEN THE SURGEON ATTEMPTED TO LOCK THE IMPLANT HEIGHT, DESPITE SEVERAL ATTEMPTS. THE ORIGINAL IMPLANT WAS REPLACED WITH A NEW IMPLANT TO COMPLETE THE SURGERY. THE ORIGINAL PRODUCT WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PROCEDURE INVOLVED DURING THE EVENT WAS A TUMOR SURGERY, DURING WHICH THE CORPECTOMY TECHNIQUE WAS APPLIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22607590 ↗MalfunctionMar 7, 2022Material Deformation; Material Deformation
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT LOCK SCREW OF CAGE BECAME CROSS THREADED DURING LOOSENING AND WOULD TORQUE OUT BEFORE BEING LOCKED INTO POSITION. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13681979 ↗MalfunctionDec 21, 2021Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING CERVICAL CORPECTOMY SPINAL THERAPY AT C4. IT WAS REPORTED THAT TOOTH ON WORM GEAR INSIDE OF CENTERPIECE BROKE OFF DURING EXPANSION. PREVENTED SURGEON FROM USING THE CAGE TO DISTRACT UNDER LOAD. THE PART CAME IN CONTACT WITH THE PATIENT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13050182 ↗MalfunctionJun 8, 2021Mechanical Jam; Mechanical Jam
INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA FIELD REPRESENTATIVE REGARDING PATIENT WITH UNKNOWN INDICATION INVOLVED IN C5-C6 CORPECTOMY WITH T2 STRATOSPHERE. IT WAS REPORTED THAT DURING PROCEDURE, TRIED TO REMOVE THE INSERTER, CAGE WOULD NOT DISENGAGE. TRIED TRANSLATING IT CEPHALAD, ROCKING IT MEDIAL-LATERAL, CEPHALAD-CAUDAL, AND 360 DEGREES. WHEN THEY PULLED THE INSERTER BACK THE CAGE CAME OUT WITH IT DUE TO THE CEPHALAD LOCKING MECHANISM NOT RELEASING FROM THE CAGE. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. PATIENT IS ALIVE AND NO INJURY. ON 2021-JUNE-07, RECEIVED ADDITIONAL INFORMATION THAT, WHEN THEY EXPANDED THE CAGE, LOCKED IT INTO PLACE, TRIED TO RELEASE THE CAGE FROM THE INSERTER AND THEY COULD NOT GET THE INSERTER TO RELEASE FROM THE IMPLANT. THERE WAS A DELAY IN PROCEDURE TIME, ESTIMATED THAT IT ADDED 50 MINUTES TO THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11963286 ↗MalfunctionMay 25, 2021Break; Activation Failure; Break; Activation Failure
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH A SPINAL PRODUCT TYPE FOR UNKNOWN INDICATION. IT WAS REPORTED THAT INTRA-OP, TEETH ON RACK/PINION PART OF IMPLANT SHEARED OFF. THIS RESULTED IN THE IMPLANT NOT EXPANDING PROPERLY AND IT WAS EXPLANTED. THE IMPLANT WAS REMOVED AND INSPECTED, AND IT WAS FOUND THAT THE EXPANSION MECHANISM WOULD NOT WORK BECAUSE IT HAD SHEARED OFF DURING THE ATTEMPT TO EXPAND. THERE WAS A DELAY OF ROUGHLY 10 MINS IN OVERALL PROCEDURE TIME. THERE WERE NO FRAGMENTS LEFT INSIDE THE PATIENT. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11875057 ↗
Other Spinal vertebral body replacement device devices
- X-CORE 7160014Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7220016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7180018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 33-54 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 5-8 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7220018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.