T2 STRATOSPHERE™ Expandable Corpectomy System 436013D (UDI 00643169663145)Recalls & reported problems

Medtronic Sofamor Danek, Inc. · CENTERPIECE 436013D 13MM D-SIZE · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169663145), received between May 25, 2021 and Jul 25, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Activation Failure1
  • Material Deformation1
  • Mechanical Jam1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2025Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER, DISTRIBUTOR) REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT DURING THE PLACEMENT OF THE CAGE, THE SCREW DRIVER SPUN FREELY AND WOULD NOT TIGHTEN WHEN THE SURGEON ATTEMPTED TO LOCK THE IMPLANT HEIGHT, DESPITE SEVERAL ATTEMPTS. THE ORIGINAL IMPLANT WAS REPLACED WITH A NEW IMPLANT TO COMPLETE THE SURGERY. THE ORIGINAL PRODUCT WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PROCEDURE INVOLVED DURING THE EVENT WAS A TUMOR SURGERY, DURING WHICH THE CORPECTOMY TECHNIQUE WAS APPLIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22607590 ↗
  • MalfunctionMar 7, 2022Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT LOCK SCREW OF CAGE BECAME CROSS THREADED DURING LOOSENING AND WOULD TORQUE OUT BEFORE BEING LOCKED INTO POSITION. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13681979 ↗
  • MalfunctionDec 21, 2021Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING CERVICAL CORPECTOMY SPINAL THERAPY AT C4. IT WAS REPORTED THAT TOOTH ON WORM GEAR INSIDE OF CENTERPIECE BROKE OFF DURING EXPANSION. PREVENTED SURGEON FROM USING THE CAGE TO DISTRACT UNDER LOAD. THE PART CAME IN CONTACT WITH THE PATIENT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13050182 ↗
  • MalfunctionJun 8, 2021Mechanical Jam; Mechanical Jam

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA FIELD REPRESENTATIVE REGARDING PATIENT WITH UNKNOWN INDICATION INVOLVED IN C5-C6 CORPECTOMY WITH T2 STRATOSPHERE. IT WAS REPORTED THAT DURING PROCEDURE, TRIED TO REMOVE THE INSERTER, CAGE WOULD NOT DISENGAGE. TRIED TRANSLATING IT CEPHALAD, ROCKING IT MEDIAL-LATERAL, CEPHALAD-CAUDAL, AND 360 DEGREES. WHEN THEY PULLED THE INSERTER BACK THE CAGE CAME OUT WITH IT DUE TO THE CEPHALAD LOCKING MECHANISM NOT RELEASING FROM THE CAGE. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. PATIENT IS ALIVE AND NO INJURY. ON 2021-JUNE-07, RECEIVED ADDITIONAL INFORMATION THAT, WHEN THEY EXPANDED THE CAGE, LOCKED IT INTO PLACE, TRIED TO RELEASE THE CAGE FROM THE INSERTER AND THEY COULD NOT GET THE INSERTER TO RELEASE FROM THE IMPLANT. THERE WAS A DELAY IN PROCEDURE TIME, ESTIMATED THAT IT ADDED 50 MINUTES TO THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11963286 ↗
  • MalfunctionMay 25, 2021Break; Activation Failure; Break; Activation Failure

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH A SPINAL PRODUCT TYPE FOR UNKNOWN INDICATION. IT WAS REPORTED THAT INTRA-OP, TEETH ON RACK/PINION PART OF IMPLANT SHEARED OFF. THIS RESULTED IN THE IMPLANT NOT EXPANDING PROPERLY AND IT WAS EXPLANTED. THE IMPLANT WAS REMOVED AND INSPECTED, AND IT WAS FOUND THAT THE EXPANSION MECHANISM WOULD NOT WORK BECAUSE IT HAD SHEARED OFF DURING THE ATTEMPT TO EXPAND. THERE WAS A DELAY OF ROUGHLY 10 MINS IN OVERALL PROCEDURE TIME. THERE WERE NO FRAGMENTS LEFT INSIDE THE PATIENT. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11875057 ↗

Other Spinal vertebral body replacement device devices

All Spinal vertebral body replacement device devices →