Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169663183), received between Mar 12, 2020 and Jul 28, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Activation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Problem1
  • Material Deformation1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2025Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING VERTEBRECTOMY S PINAL THERAPY FOR COMPRESSION FRACTURE. LEVELS IMPLANTED WERE T9/10. IT WAS REPORTED THAT THE STRATOSPHERE GRAFT WOULD NOT EXPAND. THE PRODUCT WAS EXPLANTED, AND A NEW CENTERPIECE WAS IMPLANTED. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22635376 ↗
  • MalfunctionJul 26, 2022Activation Problem; Activation Problem

    INFORMATION WAS RECEIVED FROM A HEALTH CARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A CENTERPIECE USED FOR SPINAL SURGERY. PRE-OP DIAGNOSIS WAS CANCER. PROCEDURE PERFORMED IN THE EVENT WAS CORPECTOMY. IT WAS REPORTED THAT INTRA-OPERATIVELY, CAGE CLICKING AND NOT EXPANDING PROPERLY. THE PRODUCT CAME IN CONTACT WITH PATIENT. THERE WAS A DELAY OF 3 MINUTES IN THE OVERALL PROCEDURE TIME. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15101063 ↗
  • MalfunctionMar 12, 2020Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT LATERAL CORPECTOMY SURGERY AT L2 DUE TO INFECTION. INTRA-OP, THE CAGE WOULD NOT HOLD THE DESIRED HEIGHT EVEN AFTER THE LOCKING SCREW WAS TORQUED. AS PER THE SURGEON THE LOCKING SCREW WAS STRIPPED OR THE INNER THREADS ON THE CAGE WHICH ACCEPT THE LOCKING SCREW WAS STRIPPED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9822192 ↗