T2 STRATOSPHERE™ Expandable Corpectomy System 436025ARecalls & reported problems

Medtronic Sofamor Danek, Inc. · CENTERPIECE 436025A 25MM A-SIZE · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169663213), received between Apr 14, 2020 and Jul 17, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Damaged by Another Device1
  • Material Deformation1
  • Mechanical Problem1
  • Problem with Sterilization1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 17, 2025Mechanical Problem; Problem with Sterilization; Material Deformation; Mechanical Problem; Problem with Sterilization; Material Deformation

    INFORMATION WAS RECEIVED FROM USER FACILITY (UF) VIA MANUFACTURER REPRESENTATIVE REGARDING PRODUCTS USED FOR SPINAL THERAPY. IT WAS REPORTED THAT PRIOR TO THE CASE, BOTH IMPLANTS WOULD NOT PROPERLY LOAD ONTO MULTIPLE INSERTERS, INCLUDING BOTH NON-NAVIGATED AND NAVIGATED INSERTERS. THE IMPLANTS WERE DESCRIBED AS WOBBLING AND WOULD NOT EXPAND AND CONTRACT PROPERLY. THE CENTER SCREWS ON BOTH IMPLANTS APPEARED TO BE STRIPPED, AND BOTH IMPLANTS HAD A SMALL AMOUNT OF BIO PRESENT. BOTH IMPLANTS WERE NEVER IMPLANTED. THERE WAS NO PATIENT INVOLVED REPORTED AND NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE ISSUE WAS NOT A MANUFACTURING ISSUE. THE STRIPPED SCREWS ARE A COMPONENT OF THE IMPLANT. THEY ARE NOT A SEPARATE ITEM/PRODUCT FROM THE CAGE. THERE WAS A LITTLE MANUFACTURED BONE INSIDE OF THE CAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22529196 ↗
  • MalfunctionApr 14, 2020Break; Device Damaged by Another Device; Break; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH L4 BURST FRACTURE, WHICH OCCURRED SOME TIME AGO. HE UNDERWENT L4 CORPECTOMY AND FUSION. INTRA-OP, AFTER INSERTING THE IMPLANT INTO THE POSITION, A CRACK LIKE SOUND WAS HEARD WHILE IT WAS BEING EXPANDED. THE TEETH ON THE CAGE HAD BROKEN, WHICH PREVENTED THE EXPANSION MECHANISM FROM WORKING. IT WAS DETERMINED THAT THE EXPANSION DRIVER BROKE THE TEETH OF THE IMPLANT. AS THE IMPLANT FAILED TO EXPAND, IT WAS REMOVED FROM THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9956473 ↗

Other Spinal vertebral body replacement device devices

All Spinal vertebral body replacement device devices →