T2 STRATOSPHERE™ Expandable Corpectomy System 436025ARecalls & reported problems
Medtronic Sofamor Danek, Inc. · CENTERPIECE 436025A 25MM A-SIZE · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169663213), received between Apr 14, 2020 and Jul 17, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Device Damaged by Another Device1
- Material Deformation1
- Mechanical Problem1
- Problem with Sterilization1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2025Mechanical Problem; Problem with Sterilization; Material Deformation; Mechanical Problem; Problem with Sterilization; Material Deformation
INFORMATION WAS RECEIVED FROM USER FACILITY (UF) VIA MANUFACTURER REPRESENTATIVE REGARDING PRODUCTS USED FOR SPINAL THERAPY. IT WAS REPORTED THAT PRIOR TO THE CASE, BOTH IMPLANTS WOULD NOT PROPERLY LOAD ONTO MULTIPLE INSERTERS, INCLUDING BOTH NON-NAVIGATED AND NAVIGATED INSERTERS. THE IMPLANTS WERE DESCRIBED AS WOBBLING AND WOULD NOT EXPAND AND CONTRACT PROPERLY. THE CENTER SCREWS ON BOTH IMPLANTS APPEARED TO BE STRIPPED, AND BOTH IMPLANTS HAD A SMALL AMOUNT OF BIO PRESENT. BOTH IMPLANTS WERE NEVER IMPLANTED. THERE WAS NO PATIENT INVOLVED REPORTED AND NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE ISSUE WAS NOT A MANUFACTURING ISSUE. THE STRIPPED SCREWS ARE A COMPONENT OF THE IMPLANT. THEY ARE NOT A SEPARATE ITEM/PRODUCT FROM THE CAGE. THERE WAS A LITTLE MANUFACTURED BONE INSIDE OF THE CAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22529196 ↗MalfunctionApr 14, 2020Break; Device Damaged by Another Device; Break; Device Damaged by Another Device
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH L4 BURST FRACTURE, WHICH OCCURRED SOME TIME AGO. HE UNDERWENT L4 CORPECTOMY AND FUSION. INTRA-OP, AFTER INSERTING THE IMPLANT INTO THE POSITION, A CRACK LIKE SOUND WAS HEARD WHILE IT WAS BEING EXPANDED. THE TEETH ON THE CAGE HAD BROKEN, WHICH PREVENTED THE EXPANSION MECHANISM FROM WORKING. IT WAS DETERMINED THAT THE EXPANSION DRIVER BROKE THE TEETH OF THE IMPLANT. AS THE IMPLANT FAILED TO EXPAND, IT WAS REMOVED FROM THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9956473 ↗
Other Spinal vertebral body replacement device devices
- X-CORE 7160014Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7220016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7180018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 33-54 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 5-8 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7220018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page