T2 STRATOSPHERE™ Expandable Corpectomy System 436025B (UDI 00643169663220)Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CENTERPIECE 436025B 25MM B-SIZE · Class 2 · Implant
All models of T2 STRATOSPHERE Expandable Corpectomy System (52)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169663220), received between Jul 17, 2025 and May 15, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation2
- Mechanical Problem1
- Problem with Sterilization1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2026Material Deformation
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE CENTERPIECE SET SCREW WAS STRIPPED. IT WAS ALSO REPORTED THAT AN END CAP WAS ATTACHED TO THE IMPLANT AT THE TIME OF IMPLANTING. IT WAS FURTHER REPORTED THAT AN END CAP BROKE WHILE BEING REMOVED FROM THE IMPLANT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25182065 ↗MalfunctionJul 17, 2025Mechanical Problem; Problem with Sterilization; Material Deformation; Mechanical Problem; Problem with Sterilization; Material Deformation
INFORMATION WAS RECEIVED FROM USER FACILITY (UF) VIA MANUFACTURER REPRESENTATIVE REGARDING PRODUCTS USED FOR SPINAL THERAPY. IT WAS REPORTED THAT PRIOR TO THE CASE, BOTH IMPLANTS WOULD NOT PROPERLY LOAD ONTO MULTIPLE INSERTERS, INCLUDING BOTH NON-NAVIGATED AND NAVIGATED INSERTERS. THE IMPLANTS WERE DESCRIBED AS WOBBLING AND WOULD NOT EXPAND AND CONTRACT PROPERLY. THE CENTER SCREWS ON BOTH IMPLANTS APPEARED TO BE STRIPPED, AND BOTH IMPLANTS HAD A SMALL AMOUNT OF BIO PRESENT. BOTH IMPLANTS WERE NEVER IMPLANTED. THERE WAS NO PATIENT INVOLVED REPORTED AND NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE ISSUE WAS NOT A MANUFACTURING ISSUE. THE STRIPPED SCREWS ARE A COMPONENT OF THE IMPLANT. THEY ARE NOT A SEPARATE ITEM/PRODUCT FROM THE CAGE. THERE WAS A LITTLE MANUFACTURED BONE INSIDE OF THE CAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22529215 ↗
Other Spinal vertebral body replacement device devices
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7160012Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7220018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7160014Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.