Protégé™ GPS™ SERB65-09-80-80Recalls & reported problems
Ev3, Inc · STENT SERB65-09-80-80 PROTEGE GPS V06 · Class 3 · Implant
All models of Protégé GPS (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169728998), received between Jun 8, 2020 and Jun 8, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P060001); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Fracture1
- Material Deformation1
- Premature Activation1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 8, 2020Fracture; Premature Activation; Activation, Positioning or Separation Problem; Material Deformation; Fracture; Premature Activation; Activation, Positioning or Separation Problem; Material Deformation
PHYSICIAN INTENDED TO USE A PROTEGE GPS SELF-EXPANDING STENT SYSTEM WITH A 6FR NON-MEDTRONIC SHEATH AND 0.035 NON-MEDTRONIC GUIDEWIRE DURING TREATMENT OF A SOFT TISSUE, 80MM CTO (CHRONIC TOTAL OCCLUSION-100%) IN THE PATIENT¿S MID LEFT COMMON ILIAC ARTERY OF DIAMETER 8MM. SLIGHT VESSEL TORTUOSITY IS REPORTED. NO DAMAGE WAS NOTED TO THE PRODUCT PACKAGING PRIOR TO USE. NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE PACKAGING. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. THE LESION WAS PRE-DILATED. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT AND NO RESISTANCE WAS NOTED DURING ADVANCEMENT. IT IS REPORTED DEPLOYMENT DIFFICULTIES WERE NOTED. NO DAMAGE WAS NOTED TO THE DEPLOYMENT MECHANISMS OR DEVICE HANDLE PRIOR TO DEPLOYMENT ISSUE. THE LOCK-PIN HAD BEEN REMOVED PRIOR TO ATTEMPTED DEPLOYMENT. A PARTIAL DEPLOYMENT IS REPORTED. TWO KISSING NON-MEDTRONIC BALLOON EXPANDABLE STENTS WERE DEPLOYED AND THE PROTEGE GPS STENT WAS RETRACTED TO THE BIFURCATION WITH HALF IN THE LEFT ILIAC AND HALF IN THE RIGHT. THE 2 KISSING STENTS WERE PLACED THROUGH THE INTERSTICES OF THE STENT. NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10127668 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P060001. One submission often covers many models, so check the product and lot details in each record.
- Z-0879-2015 · Dec 4, 2014 · Terminatedev3, Protg Rx Tapered Carotid Stent System, SEPX-8-6-40-135, Sterile EO, Rx only. Indicated for the treatment of patients at high risk for adverse events from carotid endarterect
- Z-0878-2015 · Dec 4, 2014 · Terminatedev3 Protege Rx Tapered Carotid Stent System, SECX-10-7-40-135, Sterile EO, Rx only. Indicated for the treatment of patients at high risk for adverse events from carotid endartere
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Absolute Pro 1012536-100Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page