CD HORIZON® SOLERA® VOYAGER™ Spinal System 55850016540Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 55850016540 5.5 VOYAGER MAS 6.5X40 · Class 2 · Implant
All models of CD HORIZON SOLERA VOYAGER Spinal System (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169768260), received between Nov 4, 2022 and Sep 29, 2026 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Deformation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation3
- Break1
- Mechanics Altered1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Mechanics Altered
Information was received from healthcare provider (HCP) via a manufacturer representative regarding a Patient having Spinal therapy. It was reported that the set screw had difficulty passing through the upper inner threads of the tulip, and the operation time was extended. Several set screws were attempted. The screw was explanted and replaced. There were no patient symptoms or complications have been reported as a result of this event. Additional information received from the manufacturer representative that the procedure involved was Transforaminal Lumbar Interbody Fusion to L4/5. Additional information received from the manufacturer representative that the delay occurred is less than 60 minutes.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928257 ↗InjuryNov 2, 2023Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING A SPINAL THERAPY. IT WAS REPORTED THAT THE SCREWS BROKE AND THE PATIENT HAD TO BE REOPERATED. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18053216 ↗InjuryJan 13, 2023Material Deformation; Material Deformation
ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE PRODUCT WILL NOT BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16156318 ↗InjuryJan 13, 2023Material Deformation; Material Deformation
ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE PRODUCT WILL NOT BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16156306 ↗InjuryNov 4, 2022Material Deformation; Material Deformation
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING PERCUTANEOUS SP INE SURGERY FOR UNKNOWN INDICATION. IT WAS REPORTED THAT ON (B)(6) 2022, AT THE CAFAM 93 CLINIC, 1-LEVEL PERCUTANEOUS SPINE SURGERY WAS STARTED ON THE PATIENT (B)(6), TWO 10-MM PAK-TYPE GUIDES WERE PASSED TO TAKE THE PEDICLES WHERE TAKING THEM IS DIFFICULT. SINCE THEY WERE NOT WELL VISUALIZED, THE NITINOL GUIDES ARE PASSED TO BE ABLE TO PLACE THE SCREWS, DR. (B)(6) TELLS ME THAT THE SCREWS TO BE PLACED MUST BE 6.5X 40MM REF 55850016540, ONCE ALL THE SCREWS HAVE BEEN PLACED, PROCEED TO MEASURE THE BAR, IT IS CUT, FOLDED AND PLACED IN THE SCREWS, IT IS VERIFIED THAT THE BAR IS INSIDE THE HEAD OF THE SCREWS, THE FIRST STOP REF 6540530 IS PLACED, WHICH ENTERS NORMALLY WHEN PLACING THE SECOND STOP, IT ENTERS WITH DIFFICULTY IS VERIFIED BY X-RAYS THE POSITION OF THE BAR, THIS IS POSITIONED MAKING THE STOP ENTERS, THE FIRST STOP IS FRACTURED, THIS FRACTURES NORMALLY WHEN FRACTURING THE SECOND, IT DOES NOT FRACTURE AND THE STOP IS TRANS-THREADED, IT IS CHANGED STOP AND THIS IS ALSO CROSS-THREADED AND DOES NOT ALLOW IT TO FRACTURE, THE SURGEON IS SUGGESTED TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15731555 ↗
Other Appliance, fixation, spinal intervertebral body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page