NuVent™ 18INFKIT (UDI 00643169770997)Recalls & reported problems

Medtronic Xomed, Inc. · INFLATOR KIT 18INFKIT BALLOON · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169770997), received between Dec 16, 2019 and Oct 7, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Separation1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 7, 2024Break; Break

    ADDITIONAL INFORMATION RECEIVED STATING THE ISSUE HAPPENED PRIOR TO SURGERY DURING PRIMING THE DEVICE, THE PLUNGER CAME OUT OF THE BACK. THEREFORE BALLOON COULD NOT BE INFLATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20382982 ↗
  • MalfunctionJun 3, 2024

    ENDOSCOPIC SINUS SURGERY- FRONTAL BALLOON SINUSOTOMY. ATTEMPTED NAVIGATED BALLOON DILATION WITHOUT SUCCESS DUE TO MEDTRONIC FRONTAL BALLOON MALFUNCTION ¿ THE BALLOON WOULD NOT FULLY INFLATE. SURGEON ATTEMPTED TROUBLESHOOTING MULTIPLE TIMES ¿ INSTRUMENT UNABLE TO COMPLETE TASK. MANUFACTURER RESPONSE FOR INSTRUMENT, ENT MANUAL SURGICAL, NUVENT ¿ (PER SITE REPORTER). NO RESPONSE AS OF YET. MANUFACTURER RESPONSE FOR INSTRUMENT, ENT MANUAL SURGICAL, NUVENT¿ (PER SITE REPORTER). NO RESPONSE AS OF YET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19446456 ↗
  • MalfunctionFeb 21, 2024Break; Break

    IT WAS REPORTED THAT THE BALLOON WAS TAKEN OUT OF THE BOX AND USED ON THE PATIENT IN THE FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS) CASE. BALLOON WAS NOT INFLATING PROPERLY AFTER USE IN THE MIDDLE OF THE PROCEDURE. WHEN THE INFLATOR WAS CHECKED TO ENSURE PROPER PRIMING, IT FELL APART IN TWO PIECES IN THE SCRUB TECH'S HANDS. THE SCRUB STATED 'THAT THEY DID WHAT THEY ALWAYS DO". THE SCRUB IS VERY FAMILIAR WITH PRODUCTS AND UTILIZES THEM FREQUENTLY IN PROCEDURE. BALLOON SEEKR DID NOT CONTRIBUTE TO THE INFLATOR EVENT. BUT DUE TO WASTED PRODUCT IT WAS CREDITED. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18745284 ↗
  • MalfunctionDec 16, 2019Material Separation; Use of Device Problem; Material Separation; Use of Device Problem

    A HEALTH CARE PROVIDER (HCP) REPORTED VIA A MANUFACTURER REPRESENTATIVE THAT DURING THE PROCEDURE, THE INFLATOR CAME APART AND USED A SYRINGE TO DEFLATE IN LEFT FRONTAL SINUS. THE PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE. THE PATIENT WAS NOT HARMED. ON FOLLOW UP, IT WAS CONFIRMED THAT THE HCP TRIED TO CONNECT THE INFLATOR BACK BUT IT DID NOT WORK AND THEY USED A SYRINGE TO DEFLATE DURING A FUNCTIONAL ENDOSCOPIC SINUS SURGERY. THERE WAS A DELAY OF LESS THAN 5 MINUTES. IT WAS NOTED THAT IT WAS THE INITIAL USE OF THE DEVICE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9475039 ↗

Other INSTRUMENT, ENT MANUAL SURGICAL devices

All INSTRUMENT, ENT MANUAL SURGICAL devices →