NIM-Response® 8253200 (UDI 00643169782440)Recalls & reported problems
Medtronic Xomed, Inc. · PATIENT INTFC 8253200 RESPONSE 3.0 ROHS · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic NIM-Response patient interface is a nerve monitoring device used during surgery. Its 150 adverse event reports include 1 injury and no deaths, and many describe no stimulation response before surgery, incorrect readings, and false negatives. Reports fell to 14 in the past 12 months from 44 before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 150
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 150 adverse event reports that list this exact device (DI 00643169782440), received between Feb 9, 2018 and Aug 4, 2026 — classified as 1 injury, 149 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (82 reports).
3 other recalls cover devices cleared under the same FDA submission (K083124); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings82
- Use of Device Problem35
- False Negative Result15
- Device Sensing Problem8
- Break7
- Connection Problem3
- No Device Output3
- Inappropriate Audible Prompt/Feedback2
Patient problems
- No Clinical Signs, Symptoms or Conditions125
- No Consequences Or Impact To Patient24
- No Known Impact Or Consequence To Patient4
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 150 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50126506 ↗MalfunctionJun 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, THE NIM PATIENT INTERFACE WAS NOT PROVIDING STIMULATION RESPONSE. SYSTEM PASSED ELECTRODE CHECK BEFORE ENTERING INTO THE MONITORING SCREEN. RETURN STIMULUS WAS DISPLAYED AS "0", THERE WAS NO AUDIBLE FEEDBACK NOR A WAVEFORM GOT DISPLAYED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25617696 ↗MalfunctionJun 15, 2026Inappropriate Audible Prompt/Feedback
IT WAS REPORTED THAT PRE-OP, THE SIGNAL INTERFERENCES WERE OBSERVED. NO TROUBLESHOOTING WAS PERFORMED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25498272 ↗MalfunctionMay 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, THE PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25340614 ↗MalfunctionMay 13, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, THE MAINFRAME HAD NOT STIMULATION RESPONSE AND PROCEDURE WAS COMPLETED WITH THE BACK UP PRODUCT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25158763 ↗MalfunctionApr 9, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, THERE WAS NO STIMULATION RESPONSE FROM PATIENT INTERFACE. THERE WAS DAMAGE TO THE STIM 1 CHANNEL AND ALL OTHER STIM CHANNELS WERE WORKING AS EXPECTED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24825157 ↗MalfunctionMar 16, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED DURING PRE-OP THAT THE DEVICE HAD NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24604092 ↗MalfunctionFeb 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, THE PI WAS NOT STIMULATING AND THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24439812 ↗MalfunctionFeb 9, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, THE PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24289085 ↗MalfunctionFeb 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT PRE-OP, THERE WAS NO STIMULATION RESPONSE. THE STIM RETURN, LOCATED UNDER THE STIMULUS SETTING ON THE MONITOR, DISPLAYED 0. THERE WAS NO STIMULUS DELIVERY TONE AND WAVEFORM DISPLAYED WHILE STIMULATION. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24256358 ↗MalfunctionDec 30, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, THE PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THE ISSUE WAS NOT RESOLVED BY TROUBLESHOOTING. BACKUP DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23919401 ↗MalfunctionDec 22, 2025False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING INTRA-OP, THERE WAS NO NERVE RESPONSE. THERE WAS NO STIMULUS DELIVERY SOUND WHEN TRYING TO STIMULATE. THE PROCEDURE WAS COMPLETED USING ANATOMICAL LANDMARKS. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23861901 ↗MalfunctionDec 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING OPERATION, THE PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23740331 ↗MalfunctionNov 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT PRIOR TO USE, PATIENT INTERFACE WAS GIVING NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23654152 ↗MalfunctionOct 31, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING PRE-OP, THE PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23433463 ↗MalfunctionOct 24, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT PATIENT INTERFACE CABLE SHORTED , WORKS INTERMITTENTLY. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23377991 ↗MalfunctionOct 21, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE PATIENT INTERFACE WAS NOT WORKING PROPERLY. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23347375 ↗MalfunctionSep 19, 2025Break; Incorrect, Inadequate or Imprecise Result or Readings; Break; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE NIM 3.0 PI HAD CONNECTOR BROKEN. THERE¿S NO STIMULATION RESPONSE ERROR BECAUSE OF CONNECTOR BROKEN. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23093137 ↗MalfunctionAug 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION THE PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22892069 ↗MalfunctionAug 11, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT PRIOR TO USE ON THE PATIENT, THE DEVICE WAS NOT WORKING PROPERLY. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22779966 ↗MalfunctionAug 7, 2025No Device Output; No Device Output
IT WAS REPORTED THAT DURING INTRA-OP, THE NIM 3.0 PATIENT INTERFACE HAD NO SIGNAL FROM ELECTRODE/STIM. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22735617 ↗MalfunctionJul 30, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION THE PATIENT INTERFACE BOX HAD NO STIMULATION RESPONSE. DURING NERVE STIMULATION ATTEMPTS, NO RESPONSE WAS DETECTED ON THE MONITOR, THE STIM RETURN SHOWED "0," NO "STIMULUS DELIVERY" TONE WAS HEARD, AND NO WAVEFORM WAS DISPLAYED. THE PROCEDURE WAS COMPLETED WITH THE OTHER EQUIPMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22665990 ↗MalfunctionJul 30, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING SETUP, THERE WAS NO STIMULATION RESPONSE FROM THE NIM DEVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22661789 ↗MalfunctionJul 17, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED DURING PRE-OP THAT THE DEVICE HAD NO STIMULATION. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22528445 ↗MalfunctionJul 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE NIM 3.0 SYSTEM PASSED THE TEST BUT DID NOT FUNCTION CORRECTLY ON THE PATIENT. THE TESTS FAILED TO TURN GREEN WHEN ACTUAL CONSUMABLES WERE CONNECTED TO THE PATIENT, DESPITE TRYING MULTIPLE SETS OF CONSUMABLES. THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22517991 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083124. One submission often covers many models, so check the product and lot details in each record.
- Z-2342-2012 · Sep 27, 2010 · TerminatedInterfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces 8253200 (Response) LABELING: Box Label: REF 8253200 x1***NIM-Response 3.0 Patient
- Z-2344-2012 · Sep 27, 2010 · TerminatedInterfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253210 (Pulse). LABELING: Box Label: REF 8253210 x1***NIM-Response 3.0 Patie
- Z-2343-2012 · Sep 27, 2010 · TerminatedInterfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253410 (Neuro). LABELING: Box Label: REF 8253410 x1***NIM-Response 3.0 Patie
Other STIMULATOR, NERVE devices
- NIM Vital NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls260 reports
- NIM TriVantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls176 reports
- NIM® 8229306Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall1 death167 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, NERVE3 recalls45 reports
- NIM Trivantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall142 reports
- NIM® 8229307Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall138 reports
- NIM TriVantage® 8229706Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls78 reports
- NIM-Response® 8253002Medtronic Xomed, Inc. · STIMULATOR, NERVE171 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.