Nerve Integrity Monitor (NIM) 3.0 8253200RFRecalls & reported problems

Medtronic Xomed, Inc. · PATIENT INTFC 8253200RF RESP 3.0 REFURB · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00643169782501), received between Nov 27, 2017 and Mar 6, 2025 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (8 reports).

3 other recalls cover devices cleared under the same FDA submission (K083124); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem8
  • Device Sensing Problem3
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Device Operates Differently Than Expected1
  • Device Stops Intermittently1
  • Electrical Shorting1
  • Failure to Sense1
  • Intermittent Continuity1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient6
  • No Known Impact Or Consequence To Patient5

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT PRE-OP, THE DEVICE HAD NO COMMUNICATION. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21536705 ↗
  • MalfunctionMay 6, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE NIM PATIENT INTERFACE WAS MALFUNCTIONING. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19246439 ↗
  • MalfunctionNov 14, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE DEVICE HAD NOT WORKING. PATIENT INTERFACE EMG TEST STAYS WITH QUESTION'S MARKS. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18130279 ↗
  • MalfunctionJun 5, 2023Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT DURING INTRA-OP, THE DEVICE WAS NOT STIMULATING. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17064827 ↗
  • MalfunctionDec 28, 2022Intermittent Loss of Power; Intermittent Loss of Power

    IT WAS REPORTED THAT DEVICE CHANNEL 1 PORT IS INTERMITTENT. THERE IS NO KNOWN PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16061802 ↗
  • MalfunctionMay 18, 2022Use of Device Problem; Use of Device Problem

    A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT THE DEVICE WAS NOT FUNCTIONING PROPERLY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14430102 ↗
  • MalfunctionDec 8, 2020Use of Device Problem; Use of Device Problem

    ON FOLLOW-UP, IT WAS REPORTED THAT THE FACILITY HAVE 2 UNITS AND THEY JUST SWAPPED THEM BEFORE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10967573 ↗
  • MalfunctionOct 14, 2020Sensing Intermittently; Failure to Sense; Device Sensing Problem; Sensing Intermittently; Failure to Sense; Device Sensing Problem

    ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT THE STIMULUS PROBE DID NOT IMMEDIATELY RESPOND. THE STIMULUS WAS TRIGGERED AGAIN AND IT ACTIVATED. THERE WAS MOMENTARY DELAY (LESS THAN 1 HOUR).

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10676350 ↗
  • MalfunctionApr 9, 2020Intermittent Continuity; Intermittent Continuity

    A HEALTH CARE PROVIDER (HCP) REPORTED THAT THE PATIENT INTERFACE WAS WORKING INTERMITTENTLY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9942161 ↗
  • MalfunctionOct 8, 2019Device Sensing Problem; Device Sensing Problem

    A HEALTHCARE PROVIDER (HCP) REPORTED THAT THE DEVICE WAS NOT SENSING STIMULATION. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9166964 ↗
  • MalfunctionJun 17, 2019Use of Device Problem; Use of Device Problem

    A HEALTH CARE PROVIDER (HCP) REPORTED THAT THE PATIENT INTERFACE WAS NOT FUNCTIONING. THERE WAS NO KNOWN PATIENT IMPACT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8703345 ↗
  • MalfunctionMar 5, 2019Use of Device Problem; Use of Device Problem

    A HEALTH CARE PROVIDER REPORTED THAT THE INTERFACE WAS NOT FUNCTIONING. THERE WERE NO PATIENT ISSUES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8391382 ↗
  • MalfunctionFeb 25, 2019Use of Device Problem; Use of Device Problem

    A HEALTH CARE PROVIDER (HCP) REPORTED VIA A MANUFACTURER REPRESENTATIVE THAT THE PATIENT INTERFACE WAS NOT WORKING. THERE WAS NO KNOWN PATIENT IMPACT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8367453 ↗
  • MalfunctionNov 19, 2018Use of Device Problem; Use of Device Problem

    A HEALTH CARE PROVIDER (HCP) REPORTED THAT THE DEVICE WAS NOT WORKING PROPERLY. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8086134 ↗
  • MalfunctionNov 9, 2018Unable to Obtain Readings; Use of Device Problem; Unable to Obtain Readings; Use of Device Problem

    A HEALTH CARE PROVIDER (HCP) REPORTED VIA DISTRIBUTOR THAT THE NERVE MONITOR WAS NOT GETTING ANY STIMULUS, NO STIMULATION RESPONSE OR NOT GETTING ANY POSITIVE/NEGATIVE SIGNALS FROM THE UNIT. THE PATIENT INTERFACE WAS STILL NOT WORKING WHICH JUST CAME FROM REPAIR. THE CASE WAS COMPLETED USING ANOTHER DEVICE. THE CONSOLE WAS WORKING FINE. THEY TESTED THE INTERFACE AND GOT NOTHING ON THE MONITOR (NO STIMULUS SOUNDS OR SIGNALS). THEY SWITCHED THE INTERFACE WITH A LOANER INTERFACE AND IT SEEMED TO OPERATE FINE. THE CONSOLE PASSED ALL SAFE TEST.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8055807 ↗
  • MalfunctionJun 1, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    THE CUSTOMER REPORTED THAT THE DEVICE WAS WORKING INTERMITTENTLY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7561975 ↗
  • MalfunctionFeb 13, 2018Device Stops Intermittently; Device Stops Intermittently

    THE CUSTOMER REPORTED THAT THE NIM SYSTEM WAS INTERMITTENTLY NOT WORKING. THE CASE WAS COMPLETED WITH A BACKUP DEVICE. THERE WAS NO PATIENT IMPACT. THIS IS REGULATORY REPORT 2 OF 2 (PATIENT INTERFACE).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7266040 ↗
  • MalfunctionNov 27, 2017Electrical Shorting; Electrical Shorting

    THE CUSTOMER REPORTED THEY WERE UNABLE TO DO ANY NERVE STIMULATION OR RECEIVE NERVE FEEDBACK INTRAOPERATIVELY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7062382 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083124. One submission often covers many models, so check the product and lot details in each record.

  • Z-2344-2012 · Sep 27, 2010 · Terminated
    Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253210 (Pulse). LABELING: Box Label: REF 8253210 x1***NIM-Response 3.0 Patie
  • Z-2342-2012 · Sep 27, 2010 · Terminated
    Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces 8253200 (Response) LABELING: Box Label: REF 8253200 x1***NIM-Response 3.0 Patient
  • Z-2343-2012 · Sep 27, 2010 · Terminated
    Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253410 (Neuro). LABELING: Box Label: REF 8253410 x1***NIM-Response 3.0 Patie

Other STIMULATOR, ELECTRICAL, EVOKED RESPONSE devices