CD HORIZON® SOLERA® VOYAGER™ Spinal System 55750014540Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 55750014540 5.5VOYAGER STMAS4.5X40 · Class 2 · Implant
All models of CD HORIZON SOLERA VOYAGER Spinal System (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169783720), received between Jan 26, 2025 and Jun 10, 2026 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 10, 2026Device Dislodged or Dislocated
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE VERTEBRAL BODIES AROUND T12 AND L1 ARE DETERIORATING. AS A RESULT, THE L1 SCREWS HAVE BACKED OUT, AND THE RODS WERE PROTRUDING. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WAS NO NEED FOR HOSPITALIZATION/PROLONGATION OF EXISTING HOSPITALIZATION, REVISION SURGERY WAS PERFORMED ON (B)(6) 2026. THE DEVICES WERE EXPLANTED AND RETURNED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25458483 ↗InjuryJun 10, 2026Device Dislodged or Dislocated
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE VERTEBRAL BODIES AROUND T12 AND L1 ARE DETERIORATING. AS A RESULT, THE L1 SCREWS HAVE BACKED OUT, AND THE RODS WERE PROTRUDING. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WAS NO NEED FOR HOSPITALIZATION/PROLONGATION OF EXISTING HOSPITALIZATION, REVISION SURGERY WAS DONE ON (B)(6). THE DEVICES WERE EXPLANTED AND RETURNED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25458361 ↗MalfunctionJan 26, 2025Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING SCREWS USED IN PLIF FOR L4/5 LCS. IT WAS REPORTED THAT WHEN THE ROD WAS INSERTED, THE ROD INTERFERED WITH THE EXTENDER TAB OF THE SCREW, CAUSING THE EXTENDER TAB TO BREAK. THE BROKEN SCREW WAS REMOVED, AND A NEW SCREW WAS INSERTED. THERE WAS A DELAY OF LESS THAN 60 MINS. SCREW CAME IN CONTACT WITH THE PATIENT. NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21237223 ↗
Other Appliance, fixation, spinal intervertebral body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026