NIM CONTACT® 8229508Recalls & reported problems
Medtronic Xomed, Inc. · EMG TUBE 8229508 NIM CONT 8MM ROHS · Class 2
All models of NIM Contact (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169789579), received between Jan 25, 2023 and Mar 14, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Inappropriate Audible Prompt/Feedback” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Apr 29, 2022Open, Classified
The firm received reports of events related to airway obstruction while using NIM Standard Reinforced EMG Endotracheal Tubes & NIM CONTACT Reinforced EMG Endotracheal Tubes.
Z-1519-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Inappropriate Audible Prompt/Feedback1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 14, 2024Incorrect, Inadequate or Imprecise Result or Readings; Inappropriate Audible Prompt/Feedback; Incorrect, Inadequate or Imprecise Result or Readings; Inappropriate Audible Prompt/Feedback
IT WAS REPORTED THAT INTRA OP DEVICE HAS UNEXPLAINED ACTIVITY ON CHANNEL 2. THERE WAS POSSIBLE SPONTANEOUS EMG AS IT HAD BIPHASIC WAVELENGTHS. TROUBLESHOOTING WAS DONE BY USER THAT USER LET 1-2ML OF AIR OUT OF THE CUFF AND SWAPPED RED AND BLUE ELECTRODES WITH EACH OTHER TO ELIMINATE A PI BOX ISSUE. THE ACTIVITY FOLLOWED THE CHANNEL WHEN SWAPPED RED AND BLUE ON PI BOX. THE ACTIVITY WENT AWAY ABOUT 5 MINUTES AFTER SWITCHING THE RED AND BLUE ELECTRODES WITH EACH OTHER ON PI BOX. USER WAS ABLE TO STIMULATE AND USE THE NIM VITAL FOR MONITORING THROUGHOUT THE CASE. AND ELECTRODE CHECK WAS PERFORMED. THERE WAS NO DELAY IN CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18905376 ↗MalfunctionJan 25, 2023Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT PATIENT WAS INTUBATED WITH ENDOTRACHEAL TUBE,WHEN THE TUBE WAS ATTACHED TO THE NERVE INTEGRITY MONITOR MACHINE, IT DID NOT WORK. THE ENDOTRACHEAL TUBE WAS FUNCTIONING PROPERLY, BUT THE NIM CAPABILITY DID NOT FUNCTION. THE TUBE WAS REPLACED WITH ANOTHER IDENTICAL DEVICE, AND THE SECOND ONE WORKED PROPERLY WITH THE NERVE INTEGRITY MONITOR MACHINE. NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16242643 ↗
Other STIMULATOR, NERVE devices
- NIM Vital NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls260 reports
- NIM TriVantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls176 reports
- NIM® 8229306Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall1 death167 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, NERVE3 recalls45 reports
- NIM Trivantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall142 reports
- NIM® 8229307Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall138 reports
- NIM TriVantage® 8229706Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls78 reports
- NIM-Response® 8253002Medtronic Xomed, Inc. · STIMULATOR, NERVE171 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.