The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169789579), received between Jan 25, 2023 and Mar 14, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Inappropriate Audible Prompt/Feedback” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 29, 2022Open, Classified

    The firm received reports of events related to airway obstruction while using NIM Standard Reinforced EMG Endotracheal Tubes & NIM CONTACT Reinforced EMG Endotracheal Tubes.

    Z-1519-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inappropriate Audible Prompt/Feedback1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 14, 2024Incorrect, Inadequate or Imprecise Result or Readings; Inappropriate Audible Prompt/Feedback; Incorrect, Inadequate or Imprecise Result or Readings; Inappropriate Audible Prompt/Feedback

    IT WAS REPORTED THAT INTRA OP DEVICE HAS UNEXPLAINED ACTIVITY ON CHANNEL 2. THERE WAS POSSIBLE SPONTANEOUS EMG AS IT HAD BIPHASIC WAVELENGTHS. TROUBLESHOOTING WAS DONE BY USER THAT USER LET 1-2ML OF AIR OUT OF THE CUFF AND SWAPPED RED AND BLUE ELECTRODES WITH EACH OTHER TO ELIMINATE A PI BOX ISSUE. THE ACTIVITY FOLLOWED THE CHANNEL WHEN SWAPPED RED AND BLUE ON PI BOX. THE ACTIVITY WENT AWAY ABOUT 5 MINUTES AFTER SWITCHING THE RED AND BLUE ELECTRODES WITH EACH OTHER ON PI BOX. USER WAS ABLE TO STIMULATE AND USE THE NIM VITAL FOR MONITORING THROUGHOUT THE CASE. AND ELECTRODE CHECK WAS PERFORMED. THERE WAS NO DELAY IN CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18905376 ↗
  • MalfunctionJan 25, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT PATIENT WAS INTUBATED WITH ENDOTRACHEAL TUBE,WHEN THE TUBE WAS ATTACHED TO THE NERVE INTEGRITY MONITOR MACHINE, IT DID NOT WORK. THE ENDOTRACHEAL TUBE WAS FUNCTIONING PROPERLY, BUT THE NIM CAPABILITY DID NOT FUNCTION. THE TUBE WAS REPLACED WITH ANOTHER IDENTICAL DEVICE, AND THE SECOND ONE WORKED PROPERLY WITH THE NERVE INTEGRITY MONITOR MACHINE. NO PATIENT IMPACT.