Evera MRI™ XT DR SureScan™ DDMB1D1 (UDI 00643169837713)Recalls & reported problems

Medtronic, Inc. · ICD-DR DDMB1D1 EVERA MRI XT US DF1 · Class 3 · Implant

All models of Evera MRI XT DR SureScan (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Evera MRI XT defibrillator is an implanted heart device. Its 825 reports include 16 deaths and 527 injuries. Many describe the device sensing heart signals poorly and delivering inappropriate shocks, and some describe infections leading to removal. 1 open recall, started in May 2023, concerns a rare chance that shock therapy delivers reduced or no energy.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
827
Deaths reported
16
Injuries reported
527
Recalls naming it
1

Summary

The FDA's MAUDE database contains 827 adverse event reports that list this exact device (DI 00643169837713), received between Apr 24, 2018 and Sep 25, 2026 — classified as 16 deaths, 527 injuries, 284 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (256 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 8 reportsDec 2023: 5 reportsJan 2024: 9 reportsFeb 2024: 10 reportsMar 2024: 9 reportsApr 2024: 11 reportsMay 2024: 5 reportsJun 2024: 6 reportsJul 2024: 11 reportsAug 2024: 5 reportsSep 2024: 6 reportsOct 2024: 10 reportsNov 2024: 8 reportsDec 2024: 4 reportsJan 2025: 5 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 6 reportsJun 2025: 5 reportsJul 2025: 11 reportsAug 2025: 7 reportsSep 2025: 7 reportsOct 2025: 7 reportsNov 2025: 4 reportsDec 2025: 3 reportsJan 2026: 8 reportsFeb 2026: 9 reportsMar 2026: 5 reportsApr 2026: 1 reportsMay 2026: 4 reportsJun 2026: 3 reportsJul 2026: 4 reportsAug 2026: 4 reportsSep 2026: 6 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 10, 2023Class IOpen, Classified

    There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

    Z-1711-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Sensing Problem256
  • Inappropriate/Inadequate Shock/Stimulation227
  • Adverse Event Without Identified Device or Use Problem185
  • Pacing Problem120
  • Electrical /Electronic Property Problem100
  • Electromagnetic Interference63
  • Over-Sensing55
  • Mechanical Problem50

Patient problems

  • Shock from Patient Lead(s)224
  • No Clinical Signs, Symptoms or Conditions183
  • Unspecified Infection153
  • No Known Impact Or Consequence To Patient93
  • No Patient Involvement42
  • No Consequences Or Impact To Patient31
  • Tachycardia21
  • Dizziness13

Recent adverse event reports

The 25 most recent of 827 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Defibrillation/Stimulation Problem; Device Sensing Problem

    Patient problems: Tachycardia

    MAUDE report 50905136 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50887102 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Medical device site infection; Wound Dehiscence

    MAUDE report 50826961 ↗
  • MalfunctionSep 17, 2026Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited oversensing resulting in misclassification of episode type and noise. The patient experienced an accident/trauma. The RA lead remains in use. Follow up with the clinic indicated that Electromagnetic Interference from the Implantable Cardioverter Defibrillator (ICD) was confirmed. No further actions were taken as a result of the event. The ICD remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50806197 ↗
  • InjurySep 11, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times and misclassification of episode type. The Implantable Cardioverter Defibrillator (ICD) de livered Anti-Tachycardia Pacing (ATP) and shocks for AT/AF with Rapid Ventricular Response (RVR) that was detected as Ventricular Tachycardia (VT). The ICD and lead remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50734012 ↗
  • MalfunctionSep 9, 2026Electromagnetic Interference

    It was reported that there was Electromagnetic Interference on the atrial channel of the Implantable Cardioverter Defibrillator (ICD). The device remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50693980 ↗
  • InjuryAug 31, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Implantable Cardioverter Defibrillator (ICD) and Right Ventricular (RV) lead were explanted. The reason for explant was not reported. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50533373 ↗
  • MalfunctionAug 26, 2026Signal Artifact/Noise; Electromagnetic Interference

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50454576 ↗
  • InjuryAug 20, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    It was reported that the Right Ventricular (RV) lead exhibited undersensing. It was noted that there was no impact to overall device function or performance. It was further reported that the Implantable Cardioverter Defibrillator (ICD) delivered Anti-Tachycardia Pacing (ATP) and a shock for a detected Ventricular Fibrillation (VF) that was suspected to be Atrial Tachycardia/Atrial Fibrillation (AT/AF). It was noted that the device was in mode switch prior to detection. The RV lead and the ICD remain in use. No patient complications have been reported as a result of this event.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50336506 ↗
  • MalfunctionAug 11, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50215264 ↗
  • InjuryJul 31, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times. Intermittent undersensing also resulted in misclassification of a detected Ventricular Tachycardia (VT) episode that was treated with Anti-Tachycardia Pacing (ATP) therapy. The RA lead remains in use. It was also reported that the patient received ATP and high voltage therapy from the Cardiac Resynchronization Therapy Defibrillator (CRT-D) for possible Atrial Tachycardia/Atrial Fibrillation (AT/AF) with fast ventricular response that the device detected as VT. There were also monitored episodes of possible AT/AF with fast ventricular response that were detected as Non-Sustained Ventricular Tachycardia. Functional undersensing in blanking/refractory on the ventricular channel also resulted in overpacing. The CRT-D remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50098384 ↗
  • InjuryJul 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Implantable Cardioverter Defibrillator (ICD) system was explanted due to infection with bacteremia. There was vegetation observed on the tricuspid valve. No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia; Fever; Unspecified Infection

    MAUDE report 50086789 ↗
  • MalfunctionJul 27, 2026Electromagnetic Interference; Pacing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50037335 ↗
  • MalfunctionJul 13, 2026Electromagnetic Interference; Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25797911 ↗
  • InjuryJun 24, 2026Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE PATIENT IS UNDER HOSPICE CARE AND THE ¿DEFIBRILLATOR KEEPS SHOCKING. THE MAGNET IS NOT WORKING, EVEN THOUGH IT IS TAPED. IT¿S BEEN GOING ON FOR PROBABLY A WEEK.¿ THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25610868 ↗
  • DeathJun 24, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PATIENT PASSED OUT AND COLLAPSED AT DIALYSIS TREATMENT AND WENT INTO CARDIAC ARREST. THE PATIENT WAS RESUSCITATED BUT THEN WENT INTO PULSELESS ELECTRICAL ACTIVITY AND EXPIRED. IT WAS UNKNOWN IF THERE WAS A PRODUCT PERFORMANCE ISSUE. CAUSES OF DEATH PROVIDED WERE CONGESTIVE HEART FAILURE AND CHRONIC KIDNEY DISEASE. NO ADDITIONAL INFORMATION WAS REPORTED OR COULD BE OBTAINED.

    Patient problems: Cardiac Arrest; Heart Failure/Congestive Heart Failure; Renal Failure; Syncope/Fainting

    MAUDE report 25608518 ↗
  • InjuryJun 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 25549566 ↗
  • InjuryMay 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO INFECTION. THE RIGHT VENTRICULAR (RV) AND RIGHT ATRIAL (RA) LEADS WERE PRESUMED TO BE TAKEN OUT OF SERVICE AS WELL; HOWEVER, THE STATUS OF THE LEADS IS UNKNOWN. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 25202968 ↗
  • MalfunctionMay 11, 2026Electromagnetic Interference

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH DIZZINESS AND NEAR SYNCOPE. THE RIGHT VENTRICULAR (RV) LEAD AND RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING WITH ELECTROMAGNETIC INTERFERENCE OBSERVED ON THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). ADDITIONALLY, THE RA LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF DETECTED EVENTS IN PROGRESS. THE ICD, RA LEAD, AND RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dizziness; No Clinical Signs, Symptoms or Conditions; Presyncope

    MAUDE report 25136537 ↗
  • InjuryMay 6, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING AND UNDERSENSING RESULTING IN POSSIBLE MISCLASSIFICATION OF EPISODE TYPE AND MAY HAVE LED TO ADDITIONAL THERAPIES DELIVERED. THE RV LEAD ALSO EXHIBITED T-WAVE OVERSENSING (TWOS),  THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND SHOCKED FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE WHICH THE DEVICE DETECTED AND TREATED AS VENTRICULAR FIBRILLATION (VF). THE LEAD  AND DEVICE REMAINS IN USED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25107404 ↗
  • InjuryMay 5, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) MAY HAVE PROVIDED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND HIGH VOLTAGE THERAPY FOR AN SUPRA VENTRICULAR TACHYCARDIA (SVT) EPISODE DETECTED AS VENTRICULAR FIBRILLATION (VF). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25083477 ↗
  • InjuryApr 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. A NEW ICD SYSTEM WAS IMPLANTED APPROXIMATELY A WEEK LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 25019386 ↗
  • InjuryMar 24, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT FOR CONGESTIVE HEART FAILURE, DIZZINESS, AND NEAR SYNCOPE. THE TRANSMISSION REPORT WAS REVIEWED AND IT WAS NOTED THAT THE PATIENT RECEIVED POSSIBLY INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND HIGH-VOLTAGE THERAPY FOR ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) WITH RAPID VENTRICULAR RESPONSE THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) CLASSIFIED AS VENTRICULAR FIBRILLATION (VF). THE DEVICE WAS NOTED TO BE AT END OF SERVICE (EOS); THEREFORE, THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dizziness; Heart Failure/Congestive Heart Failure; Presyncope; Shock from Patient Lead(s)

    MAUDE report 24681118 ↗
  • InjuryMar 20, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED POSSIBLE INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND SHOCKS FOR VENTRICULAR TACHYCARDIA (VT) THAT WAS SUSPECTED TO BE SUPRA VENTRICULAR TACHYCARDIA (SVT). IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING OF ATRIAL FIBRILLATION (AF) RESULTING IN INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP). REPROGRAMMING WAS DONE ON THE RA LEAD. THE ICD AND RA LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24654046 ↗
  • InjuryMar 19, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POTENTIAL INAPPROPRIATE THERAPY DURING TREATED VENTRICULAR FIBRILLATION (VF), WHERE THERE WAS ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION IN PROGRESS AT THE TIME OF THE EPISODES. IN ADDITION, THERE WAS UNDERSENSING OBSERVED ONTHE RIGHT VENTRICULAR (RV) AND RIGHT ATRIAL (RA) LEADS. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)