Visa AF MRI™ VR SureScan™ DVFB1D1Recalls & reported problems

Medtronic, Inc. · ICD-VR DVFB1D1 VISIA AF MRI US DF1 · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Visia AF MRI defibrillator has 269 reports, including 4 deaths and 217 injuries. Many describe sensing problems or shocks given when not needed, and some describe infections that led to removal. One open recall, which began in May 2023, concerns rare cases where a shock may deliver little or no energy. Reports dropped sharply over the last 12 months, to 13 from 59 the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
267
Deaths reported
4
Injuries reported
215
Recalls naming it
1

Summary

The FDA's MAUDE database contains 267 adverse event reports that list this exact device (DI 00643169837799), received between May 10, 2018 and Jul 2, 2026 — classified as 4 deaths, 215 injuries, 48 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (107 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 5 reportsApr 2024: 6 reportsMay 2024: 1 reportsJun 2024: 5 reportsJul 2024: 2 reportsAug 2024: 3 reportsSep 2024: 3 reportsOct 2024: 4 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 3 reportsMay 2025: 2 reportsJun 2025: 3 reportsJul 2025: 4 reportsAug 2025: 13 reportsSep 2025: 25 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202625
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 10, 2023Class IOpen, Classified

    There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

    Z-1803-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Sensing Problem107
  • Inappropriate/Inadequate Shock/Stimulation102
  • Adverse Event Without Identified Device or Use Problem71
  • Pacing Problem21
  • Mechanical Problem16
  • Over-Sensing16
  • Electromagnetic Interference15
  • Signal Artifact/Noise13

Patient problems

  • Shock from Patient Lead(s)101
  • Unspecified Infection56
  • No Clinical Signs, Symptoms or Conditions39
  • No Known Impact Or Consequence To Patient21
  • Tachycardia7
  • Bradycardia6
  • No Consequences Or Impact To Patient5
  • No Patient Involvement5

Recent adverse event reports

The 25 most recent of 267 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 2, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. GROWTH WAS NOTED ON THE SUPERIOR VENA CAVA (SVC) COIL OF THE RIGHT VENTRICULAR (RV) LEAD. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Medical device site infection

    MAUDE report 25695190 ↗
  • DeathApr 21, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE PATIENT REPRESENTATIVE THAT THE PATIENT HAD PASSED AWAY. THE REPRESENTATIVE WANTED TO KNOW IF THE DEVICE WAS WORKING APPROPRIATELY AT TIME OF PASSING, AND STATED, "IT DIDN'T LOOK LIKE IT WAS WORKING". IT WAS ALSO STATED THAT THE PATIENT UNDERWENT A FISTULA PROCEDURE ONE OR TWO DAYS PRIOR TO PASSING AWAY. NO FURTHER DETAILS WERE OBTAINED THROUGH FOLLOW-UP.

    Patient problems: Insufficient Information

    MAUDE report 24944937 ↗
  • MalfunctionApr 17, 2026Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS FALSELY DETECTING ATRIAL FIBRILLATION (AF) EPISODES. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24911480 ↗
  • InjuryMar 26, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED A VENTRICULAR FIBRILLATION (VF) EPISODE AND DURING THE DETECTION AND POST THERAPY SENSING PERIOD, A DEVICE DETECTION RELATED OBSERVATION WAS NOTED, CHARACTERIZED BY ONE BEAT OF VENTRICULAR UNDERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD IMMEDIATELY FOLLOWING SHOCK DELIVERY. THE EPISODE WAS DETECTED AND DELIVERED ANTITACHYCARDIA PACING FOLLOWED BY A SHOCK. THE RV LEAD AND ICD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24701629 ↗
  • InjuryMar 19, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION AND POCKET EROSION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pocket Erosion; Unspecified Infection

    MAUDE report 24642841 ↗
  • InjuryFeb 3, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED SUSPECT INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND HIGH-VOLTAGE THERAPY FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) CLASSIFIED AS FAST VENTRICULAR TACHYCARDIA (FVT). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24251321 ↗
  • InjuryJan 16, 2026Failure to Deliver Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WITHHELD THERAPY FOR VENTRICULAR TACHYCARDIA (VT) THAT WAS DETECTED AS SUPRA VENTRICULAR TACHYCARDIA (SVT). THE PATIENT WAS EXTERNALLY CARDIOVERTED. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.;

    Patient problems: Tachycardia

    MAUDE report 24100828 ↗
  • InjuryDec 18, 2025Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS FURTHER REPORTED THAT THE ICD SYSTEM WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Infection

    MAUDE report 23843482 ↗
  • InjuryNov 17, 2025Failure to Deliver Shock/Stimulation; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) POSSIBLY DETECTED VENTRICULAR TACHYCARDIA (VT) AS SUPRA VENTRICULAR TACHYCARDIA (SVT). THE PATIENT REQUIRED AN EXTERNAL DEFIBRILLATION. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Tachycardia

    MAUDE report 23572253 ↗
  • InjuryNov 11, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE HIGH VOLTAGE THERAPY AND INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED AS VENTRICULAR TACHYCARDIA (VT). IT WAS ALSO REPORTED THAT THERE WAS POSSIBLE NOISE ON THE RIGHT VENTRICULAR (RV) LEAD. THE ICD AND RV LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 23524641 ↗
  • InjuryNov 4, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED ANTI-TACHYCARDIA PACING (ATP) AND HIGH VOLTAGE THERAPY FROM THE SINGLE CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) FOR POSSIBLE SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT THE DEVICE CLASSIFIED AS VENTRICULAR TACHYCARDIA (VT). THE DEVICE DISCRIMINATOR STOPPED WITHHOLDING THERAPY, POSSIBLY DUE TO NOISE ON THE RIGHT VENTRICULAR CHANNEL. THE ICD AND RIGHT VENTRICULAR (RV) LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 23465042 ↗
  • InjuryOct 22, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) POSSIBLY DELIVERED AN INAPPROPRIATE SHOCK AND ANTI-TACHYCARDIA PACING (ATP) FOR SUPRA VENTRICULAR TACHYCARDIA (SVT). THE ICD DETECTED THE EPISODE AS VENTRICULAR FIBRILLATION (VF). THE PATIENT EXPERIENCED PALPITATIONS. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Arrhythmia; Shock from Patient Lead(s)

    MAUDE report 23363596 ↗
  • InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION OF THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THE ORGANISM WAS CANDIDA AURIS WHICH WAS RESISTANT TO DRUG THERAPY. THE SOURCE OF THE INFECTION WAS UNKNOWN, AND IT WAS NOTED THAT THE PATIENT IS ON DIALYSIS. THE ICD WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Fungal Infection

    MAUDE report 23315880 ↗
  • InjurySep 18, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR HIGH THRESHOLDS. APPROXIMATELY THREE YEARS LATER, THE PATIENT RECEIVED ANTI-TACHYCARDIA PACING (ATP) AND A SHOCK FOR AN EPISODE DETECTED IN THE VENTRICULAR FIBRILLATION (VF) ZONE THAT WAS SUSPECTED TO BE SUPRA VENTRICULAR TACHYCARDIA (SVT). THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND RV LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 23090912 ↗
  • InjurySep 15, 2025Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED BY THE PATIENT THAT THEY WERE SHOCKED DURING THE NIGHT, BUT DID NOT THINK THEY SHOULD HAVE BEEN. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO ADDITIONAL INFORMATION WAS OBTAINED THROUGH FOLLOW-UP WITH THE CLINIC. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 23057077 ↗
  • InjurySep 4, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR) DETECTED IN THE VENTRICULAR FIBRILLATION (VF) ZONE. THE DATA ALSO SHOWED RIGHT VENTRICULAR (RV) LEAD NOISE HAD IMPACTED EPISODE DISCRIMINATION. THE ICD AND RV LEAD WERE LATER EXPLANTED AND RETURNED TO THE MANUFACTURER AFTER THE PATIENT PASSED AWAY ONE MONTH LATER UNRELATED TO THE DEVICE SYSTEM.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22962721 ↗
  • InjurySep 3, 2025Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation; Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED DUE TO ELECTROMAGNETIC INTERFERENCE (EMI). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22957192 ↗
  • InjurySep 3, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INAPPROPRIATE THERAPY BY THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) FOR THE DETECTION OF SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT WAS CLASSIFIED AS VENTRICULAR TACHYCARDIA (VT). IT WAS NOTED THAT THERAPIES WERE INITIALLY WITHHELD BY THE DISCRIMINATOR FEATURE THAT IS DESIGNED TO WITHHOLD INAPPROPRIATE VENTRICULAR DETECTION IF THE RHYTHM DISPLAYS CHARACTERISTICS OF AN SVT ORIGIN. IT WAS ALSO NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED POSSIBLE INTERMITTENT OVERSENSING ON STORED VENTRICULAR ELECTROGRAMS (EGM). THE DEVICE AND LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22954510 ↗
  • InjurySep 3, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) DETECTED IN THE VENTRICULAR FIBRILLATION (VF) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22952270 ↗
  • InjurySep 2, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) DETECTED IN THE VENTRICULAR FIBRILLATION (VF) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22946361 ↗
  • InjurySep 2, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22945927 ↗
  • InjurySep 2, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR) DETECTED IN THE VENTRICULAR FIBRILLATION (VF) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22945905 ↗
  • InjurySep 2, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) DETECTED IN THE VENTRICULAR FIBRILLATION (VF) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22945861 ↗
  • InjurySep 2, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DEVICE DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR) DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22945653 ↗
  • InjurySep 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 22945426 ↗

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →