Sprinter™ SPR2012WRecalls & reported problems
Medtronic, Inc. · BALLOON SPR2012W SPRINTER OTW 2.0X12MM · Class 2
All models of Sprinter (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169850385), received between Feb 7, 2022 and Feb 21, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Difficult to Remove1
- Material Deformation1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 21, 2024Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
AN ATTEMPT WAS MADE TO USE ONE SPRINTER OTW BALLOON CATHETER TO TREAT A SEVERELY TORTUOUS, SEVERELY CALCIFIED LESION WITH 90% STENOSIS IN THE PROXIMAL LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED WHEN ATTEMPTING TO REMOVE THE BALLOON FOLLOWING BALLOON INFLATION. THE PATIENT IS ALIVE WITH NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18747462 ↗MalfunctionMar 21, 2022Use of Device Problem; Use of Device Problem
ONE SPRINTER OTW BALLOON CATHETER WAS USED. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING. THE DEVICE WAS REMOVED FROM THE HOOP WITH NO ISSUES NOTED. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE, EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT THE DEVICE WAS EXPIRED WHEN IT WAS USED IN THE PATIENT. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13838862 ↗MalfunctionFeb 7, 2022Material Deformation; Material Deformation
ONE COUGAR GUIDEWIRE (PLI10) AND ONE SPRINTER OTW BALLOON (PLI20) WERE ATTEMPTED TO BE USED DURING A PROCEDURE TO TREAT A MILDLY CALCIFIED LESION IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THERE WERE NO ISSUES NOTED TO THE DEVICE PACKAGING. THE DEVICES WERE REMOVED FROM THE PACKAGING PER IFU, WITH NO ISSUES NOTED. THE DEVICES WERE INSPECTED PRIOR TO USE WITH NO ISSUES NOTED. THERE WERE NO DIFFICULTIES NOTED WHEN REMOVING THE PROTECTIVE SHEATH OR THE PACKAGING STYLETTE FROM THE SPRINTER BALLOON. THE DEVICES WERE PREPPED PER IFU WITH NO ISSUES NOTED. THE COUGAR GUIDEWIRE AND THE GUIDEWIRE LUMEN IN THE SPRINTER BALLOON WERE FLUSHED BEFORE USE. THE LESION WAS PRE-DILATED. THE DEVICES DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICES BUT EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT DIFFICULTIES WERE NOTED WHEN LOADING THE BALLOON ONTO THE GUIDEWIRE. IT WAS STATED THAT THE BALLOON SEEMED TO GET STUCK ON THE WIRE. THE EVENT OCCURRED WHILE ATTEMPTING TO DELIVER THE SPRINTER BALLOON TO THE LESION SITE. THE SPRINTER BALLOON DID EXIT THE TIP OF THE GUIDE CATHETER. THE GUIDEWIRE WAS FRONT LOADED THROUGH THE WIRE ENTRY PORT. THE GUID
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13471857 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗