NA 8X64R4Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK 8X64R4 10PK SUPP TUB · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169862852), received between Mar 22, 2023 and Mar 22, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
10 other recalls cover devices cleared under the same FDA submission (K800178); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak1
- Material Separation1
Patient problems
- Hemorrhage/Blood Loss/Bleeding1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 22, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT A CONNECTOR DETACHED FROM THE TUBING DURING USE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE LARGE-BORE FEMALE LUER CONNECTOR DETACHED FROM THE TUBING WHILE THE PATIENT WAS CONNECTED. THE DETACHMENT CAUSED PATIENT BLOOD LOSS ESTIMATED BY THE CLINICIAN OF NO MORE THAN 50CC. THE CUSTOMER WAS ABLE TO RE-ATTACH THE CONNECTOR TO THE TUBING AND SECURE IT WITH A ZIP TIE. THE DETACHMENT AND BLOOD LOSS DID NOT RESULT IN ANY ADVERSE IMPACT TO THE PATIENT. NO CHANGE OUT WAS REQUIRED AND SCD THERAPY RESUMED WITH THE EXISTING CARTRIDGE AND TUBING SET. THIS INCIDENT WAS ISOLATED AS SEVERAL BLOOD TUBING SETS FROM THE SAME KIT OF 6 SETS WERE USED PRIOR TO THE MALFUNCTION WITH NO ISSUES.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 16591350 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K800178. One submission often covers many models, so check the product and lot details in each record.
- Z-2736-2015 · Aug 18, 2015 · TerminatedMEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9, 8C72R4, BB5J59R19, 6B65R3
- Z-1000-2014 · Jan 30, 2014 · TerminatedMedtronic Intersept Custom Tubing Pack which contain Select 3D or Select CAP Arterial Cannulae. Model 3Y90R3, the following model is only approved in Europe C300901B. Sterilized
- Z-2104-2013 · Jul 17, 2013 · TerminatedMedtronic Custom Perfusion System. Model Numbers: 1D80R8, 1E34R1, 1F75R1, 1G49R2, 1G75R1, 1P07R8, 1P91R4, 5B15R11, and 5P51R9. Sterilized by Ethylene Oxide, Do no Reuse. Pr
- Z-0618-2013 · Nov 19, 2012 · TerminatedMedtronic Intersept Custom Tubing Pack with or without coating ( Carmeda BioActive Surfact, Carmeda BioActive Surfact and Trillium Biosurface,or Balance Biosurface) with the follow
- Z-0814-2012 · Dec 14, 2011 · TerminatedCustom Packs/ Kerkrade. Contain Medtronic Venous Cannula, Sterile EO. Model #'s C330201C, C330201D, C330201E, C330206B, C342603A, C350106A, C390101D, C390203D, C390406C, C390409
- Z-1424-2009 · Feb 25, 2009 · TerminatedMedtronic, Transfer Spike, Blood Processing Accessory, REF BT945, Sequestration Kits
- Z-1227-2007 · Aug 1, 2007 · TerminatedMedtronic Intersept 3608 Sterile tubing (8') 1/2 x 3/32 in. Quantity 10, Lot 7626621, Use by 2009-07-31. Manufacturer: Medtronic, Inc. Minneapolis, MN 55432-5604 USA. Diffe
- Z-0583-2007 · Jan 2, 2007 · TerminatedMedtronic 2T10R1 Intersept Custom Tubing Pack. Pack 2T10R1 consists of two PVC lines; one of them is 132 inch in length and is capped. The second line is 102 inch in length and
- Z-0340-2007 · Sep 13, 2006 · TerminatedMedtronic Intersept Custom Tubing Pack. Do Not Reuse. The design and components of this product were specified by the user from an array of components and designs offered by the
- Z-0808-05 · Mar 25, 2005 · TerminatedMedtronic Custom Packs and Total System Packs. Pack can contain a number of different components specified by the customer and individualized for their bypass circuit set up requ
Other TUBING, PUMP, CARDIOPULMONARY BYPASS devices
- NA BB9L72R1Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS3 recalls0 reports
- NA BB8M16R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls1 report
- NA BB8D40R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8T77R2Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8L58R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB7P46R4Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB7P48R4Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8B70R7Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.