Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169862852), received between Mar 22, 2023 and Mar 22, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

10 other recalls cover devices cleared under the same FDA submission (K800178); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Material Separation1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 22, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT A CONNECTOR DETACHED FROM THE TUBING DURING USE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE LARGE-BORE FEMALE LUER CONNECTOR DETACHED FROM THE TUBING WHILE THE PATIENT WAS CONNECTED. THE DETACHMENT CAUSED PATIENT BLOOD LOSS ESTIMATED BY THE CLINICIAN OF NO MORE THAN 50CC. THE CUSTOMER WAS ABLE TO RE-ATTACH THE CONNECTOR TO THE TUBING AND SECURE IT WITH A ZIP TIE. THE DETACHMENT AND BLOOD LOSS DID NOT RESULT IN ANY ADVERSE IMPACT TO THE PATIENT. NO CHANGE OUT WAS REQUIRED AND SCD THERAPY RESUMED WITH THE EXISTING CARTRIDGE AND TUBING SET. THIS INCIDENT WAS ISOLATED AS SEVERAL BLOOD TUBING SETS FROM THE SAME KIT OF 6 SETS WERE USED PRIOR TO THE MALFUNCTION WITH NO ISSUES.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16591350 ↗

FDA clearance and classification

Cleared through 510(k) K800178 on Feb 21, 1980 (decision: Substantially Equivalent). Submitted by Extracorporeal Medical Specialities, Inc. as “Extracorp.'S Intersept Cus. Tubing Pk.”. The FDA review took 24 days. FDA record ↗

The FDA regulates product code DWE (tubing, pump, cardiopulmonary bypass) under 21 CFR 870.4390 as a Class II device. 21 CFR 870.4390 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K800178. One submission often covers many models, so check the product and lot details in each record.

  • Z-2736-2015 · Aug 18, 2015 · Terminated
    MEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9, 8C72R4, BB5J59R19, 6B65R3
  • Z-1000-2014 · Jan 30, 2014 · Terminated
    Medtronic Intersept Custom Tubing Pack which contain Select 3D or Select CAP Arterial Cannulae. Model 3Y90R3, the following model is only approved in Europe C300901B. Sterilized
  • Z-2104-2013 · Jul 17, 2013 · Terminated
    Medtronic Custom Perfusion System. Model Numbers: 1D80R8, 1E34R1, 1F75R1, 1G49R2, 1G75R1, 1P07R8, 1P91R4, 5B15R11, and 5P51R9. Sterilized by Ethylene Oxide, Do no Reuse. Pr
  • Z-0618-2013 · Nov 19, 2012 · Terminated
    Medtronic Intersept Custom Tubing Pack with or without coating ( Carmeda BioActive Surfact, Carmeda BioActive Surfact and Trillium Biosurface,or Balance Biosurface) with the follow
  • Z-0814-2012 · Dec 14, 2011 · Terminated
    Custom Packs/ Kerkrade. Contain Medtronic Venous Cannula, Sterile EO. Model #'s C330201C, C330201D, C330201E, C330206B, C342603A, C350106A, C390101D, C390203D, C390406C, C390409
  • Z-1424-2009 · Feb 25, 2009 · Terminated
    Medtronic, Transfer Spike, Blood Processing Accessory, REF BT945, Sequestration Kits
  • Z-1227-2007 · Aug 1, 2007 · Terminated
    Medtronic Intersept 3608 Sterile tubing (8') 1/2 x 3/32 in. Quantity 10, Lot 7626621, Use by 2009-07-31. Manufacturer: Medtronic, Inc. Minneapolis, MN 55432-5604 USA. Diffe
  • Z-0583-2007 · Jan 2, 2007 · Terminated
    Medtronic 2T10R1 Intersept Custom Tubing Pack. Pack 2T10R1 consists of two PVC lines; one of them is 132 inch in length and is capped. The second line is 102 inch in length and
  • Z-0340-2007 · Sep 13, 2006 · Terminated
    Medtronic Intersept Custom Tubing Pack. Do Not Reuse. The design and components of this product were specified by the user from an array of components and designs offered by the
  • Z-0808-05 · Mar 25, 2005 · Terminated
    Medtronic Custom Packs and Total System Packs. Pack can contain a number of different components specified by the customer and individualized for their bypass circuit set up requ

Common questions about the NA 8X64R4

Has the NA 8X64R4 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00643169862852), based on the FDA recall file dated Nov 27, 2026. However, 10 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the NA 8X64R4?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Mar 22, 2023. That report lists device problems “Fluid/Blood Leak” (1) and “Material Separation” (1); patient problem “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 8X64R4?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 8X64R4 FDA approved?

It was cleared through 510(k) K800178 on Feb 21, 1980. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NA 8X64R4, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NA, model 8X64R4). Its record shows it as in commercial distribution.