Skip to content

Evera™ S VR DVBC3D4 (UDI 00643169929876)Recalls & reported problems

Medtronic, Inc. · ICD-VR DVBC3D4 EVERA S US DF4 · Class 3 · Implant

All models of Evera S VR (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169929876), received between Aug 16, 2018 and Jun 28, 2022 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Electrical /Electronic Property Problem1
  • Mechanical Problem1
  • Premature Discharge of Battery1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS REMOVED FOR INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 14844338 ↗
  • MalfunctionDec 13, 2018Premature Discharge of Battery; Electrical /Electronic Property Problem; Premature Discharge of Battery; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAS PREMATURE BATTERY DEPLETION. THE BATTERY LONGEVITY ESTIMATE WAS LESS THAN EXPECTED. THE DEVICE CONTINUES TO BE MONITORED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8162180 ↗
  • InjuryAug 16, 2018Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE PATIENTS RIGHT VENTRICULAR (RV) LEAD HAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) WITH INCREASED SENSING INTEGRITY COUNTER (SIC), HIGH RATE NON-SUSTAINED EPISODES, NOISE, HIGH THRESHOLDS, LOW DEFIB IMPEDANCE, AND NON-CAPTURE. IT WAS THOUGHT THAT THERE COULD BE A CONNECTION ISSUE IN WHICH THE LEAD HAD NOT BEEN FULLY INSERTED INTO THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HEADER OR THE LEAD IS DISLODGED. A REVISION WAS COMPLETED AND THE DEVICE WAS EXTRACTED AND REPLACED, WHILE THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7789724 ↗

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →

Sources and updates