Skip to content

Integrated Power Console 1898430Recalls & reported problems

Medtronic Xomed, Inc. · FOOT SWITCH 1898430 ROHS · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00643169931398), received between Dec 19, 2022 and Dec 26, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Device Remains Activated” (5 reports).

1 other recall cover devices cleared under the same FDA submission (K081277); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Remains Activated5
  • Self-Activation or Keying2
  • Intermittent Continuity1
  • Unintended Power Up1
  • Unintended System Motion1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information1
  • Perforation1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 26, 2025Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT THE FOOTSWITCH WAS RUNNING WITHOUT STEPPING ON THE PEDAL. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 23894813 ↗
  • MalfunctionJun 10, 2025Unintended Power Up; Unintended Power Up

    IT WAS REPORTED THAT PRE-OP, FOOTSWITCH WAS ACTIVATING AUTOMATICALLY WITHOUT PRESSING IT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22178534 ↗
  • MalfunctionApr 4, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT THE DEVICE RUNS CONTINUOUSLY. THERE WAS NO PATIENT IMPACT IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21770544 ↗
  • InjuryOct 29, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING BRAINSURGERY, WHEN DRILLING IN THE SKULL BONE, ITEMS DID NOT STOP DRILLING WHEN SKULL BONE WAS BREACHED. SO DURA TEAR IN ALL THE 3 HOLES. THE SURGERY WAS EXTENDED FOR 30 MINUTES. THE PROCEDURE WAS COMPLETED WITH THE REPORTED PRODUCT. USER FIXED THE DURA WITH SPECIFIC PRODUCTS.

    Patient problems: Perforation

    MAUDE report 20558736 ↗
  • MalfunctionOct 29, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT INTRA OP, DEVICE STARTED TO RUN ON OWN AFTER 20MIN OF BEING ATTACHED TO IPC POWER CONSOLE. IPC FOOT PEDAL ATTACHED BUT NOT ENGAGED BY ANYONE/ANYTHING. NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20554913 ↗
  • MalfunctionJul 4, 2024Unintended System Motion; Unintended System Motion

    IT WAS REPORTED THAT THE FOOT SWITCH ACTIVATES ITSELF AFTER A FEW MINUTES. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19674947 ↗
  • MalfunctionJul 1, 2024Self-Activation or Keying; Use of Device Problem; Self-Activation or Keying; Use of Device Problem

    IT WAS REPORTED THAT POST-OP, UNTIMELY ACTIVATION OF IPC, PEDAL WAS INVOLVED. THE FOOT PEDAL WAS NOT WORKING PROPERLY. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643227 ↗
  • MalfunctionSep 12, 2023Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE THE DEVICE KEEPS SPINNING THE HANDPIECE WILL NOT STOP IT. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17729202 ↗
  • MalfunctionDec 19, 2022Intermittent Continuity; Device Remains Activated; Intermittent Continuity; Device Remains Activated

    IT WAS REPORTED BY HEALTH CARE PROFESSIONAL THAT INTRA-OP WHEN THIS FOOTSWITCH IS USED THE DRILL RUNS WITHOUT THE FOOTSWITCH BEING P RESSED(INTERMITTENT), A FOOTSWITCH FROM ANOTHER AVAILABLE SYSTEM WAS BORROWED AND THE SYSTEM WORKED CORRECTLY WITHOUT ANY ISSUES. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16006469 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081277. One submission often covers many models, so check the product and lot details in each record.

  • Z-0961-2022 · Dec 15, 2021 · Terminated
    Software 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND E

Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices

All DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page