Skip to content

EverCross™ AB35W03080135Recalls & reported problems

Ev3, Inc · PTA AB35W03080135 EVERCROSS 035 V08 · Class 2

All models of EverCross (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169973299), received between Jan 11, 2024 and Jan 11, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 11, 2024Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO USE A EVERCROSS TO TREAT A CALCIFIED LESION IN THE MID SUPERFICIAL FEMORAL ARTERY. DEGREE OF TORTUOSITY WAS LITTLE. DEGREE OF CALCIFICATION WAS SEVERE. LESION STENOSIS WAS CTO (CHRONIC TOTAL OCCLUSION 100%). EMBOLIC PROTECTION WAS NOT USED. A 6-0 SPIDER WAS USED. A MERIT INFLATION DEVICE WAS USED. IFU WAS FOLLOWED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. THERE WAS A CIRCUMFERENTIAL/RADIAL BURST DURING BALLOON INFLATION AT 3ATMS ON THE FIRST INFLATION. ALL FRAGMENTS OF THE BALLOON WERE RETRIEVED. IT WAS REPORTED THAT BALLOON DID NOT CROSS EASILY OVER SPIDER WIRE BUT WAS ABLE TO CROSS LESION. CROSSED LESION, PHYSICIAN BEGAN TO INFLATE AND NOTICED BALLOON WAS BURST. PROCEDURE COMPLETED BY SWAPPING BALLOON OUT FOR NANOCROSS OF THE SAME SIZE AND INFLATED WITH NO ISSUES. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18493344 ↗

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page