EverCross™ AB35W05200135 (UDI 00643169973794)Recalls & reported problems
Ev3, Inc · PTA AB35W05200135 EVERCROSS 035 V08 · Class 2
All models of EverCross (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169973794), received between Apr 7, 2021 and Apr 23, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Deflation Problem1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2024Deflation Problem; Deflation Problem
PHYSICIAN INTENDED TO USE AN EVERCROSS 035 PTA BALLOON WITH A 6X45 NON-MEDTRONIC SHEATH AND AN .035 NON-MEDTRONIC GUIDEWIRE DURING T REATMENT OF A 200MM PLAQUE LESION IN THE PATIENT'S LEFT MID DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). MINIMAL VESSEL TORTUOSITY WAS REPORTED. LESION EXHIBITED 50% STENOSIS. ARTERY DIAMETER REPORTED AS 6MM. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE IFU (INSTRUCTION FOR USE) WAS FOLLOWED DURING PREPARATION, PROCEDURE, POST-PROCEDURE. EMBOLIC PROTECTION WAS NOT USED. THE VESSEL WAS PRE-DILATED WITH A 3MM BALLOON WITH NO COMPLICATIONS. THE VESSEL WAS NOT POST-DILATED. A NON-MEDTRONIC INFLATION DEVICE WAS USED FOR BALLOON INFLATION. 50/50 CONTRAST INFLATION FLUID WAS USED. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. DEFLATION DIFFICULTIES WERE REPORTED. THERE WERE NO ISSUES NOTED DURING INFLATION. THERE WAS NO DAMAGE NOTED TO THE BALLOON CATHETER. WHEN THE PHYSICIAN TRIED TO DEFLATE THE EVERCROSS BALLOON, IT WOULD NOT CO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19163881 ↗MalfunctionApr 7, 2021Leak/Splash; Leak/Splash
PHYSICIAN ATTEMPTED TO USE AN EVERCROSS PTA BALLOON DURING TREATMENT OF A CALCIFIED LESION IN THE PATIENT¿S DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). MODERATE VESSEL CALCIFICATION IS REPORTED. THERE WAS NO DAMAGE TO THE PRODUCT PACKAGING. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE PACKAGING. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. EMBOLIC PROTECTION WAS NOT USED. A SYRINGE WAS USED FOR BALLOON INFLATION. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT. A BURST/LEAK/RUPTURE ON THE CATHETER WAS REPORTED AS 10ATM ON FIRST INFLATION. DURING THE INFLATION IT WAS NOTED THE PRESSURE IN THE INFLATION SYRINGE BEGAN TO DROP SLOWLY BUT THE BALLOON WAS STILL INFLATED. UPON DEFLATION THERE WAS A SMALL AMOUNT OF BLOOD RETURN INSIDE THE SYRINGE. THE BALLOON WAS PLACED ON THE BACK TABLE AND INFLATED. DURING INFLATION A LEAK WAS NOTICE IN THE CATHETER JUST BEFORE THE PROXIMAL END OF THE BALLOON. THE PROCEDURE WAS COMPLETED WITH PTA BALLOON ANGIOPLASTY AND DCB. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11626680 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page