EverCross™ AB35W07040135 (UDI 00643169974098)Recalls & reported problems
Ev3, Inc · PTA AB35W07040135 EVERCROSS 035 V08 · Class 2
All models of EverCross (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169974098), received between Nov 5, 2020 and May 23, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel2
- Detachment of Device or Device Component1
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 23, 2023Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component; Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO USE AN EVERCROSS TO TREAT A CALCIFIED LESION IN THE RIGHT MID DISTAL ILIAC ARTERY - COMMON. DEGREE OF TORTUOSITY WAS LITTLE. DEGREE OF CALCIFICATION WAS SEVERE. LESION STENOSIS WAS CTO (CHRONIC TOTAL OCCLUSION-100%). A 6 FR NON-MEDTRONIC SHEATH WAS USED. A NON-MEDTRONIC 035 GUIDEWIRE WAS USED. EMBOLIC PROTECTION WAS NOT USED. BALLOON WAS INFLATED WITH A NON-MEDTRONIC INFLATION DEVICE. SALINE/CONTRAST MIX INFLATION FLUID WAS USED. DEVICE WAS PREPPED PER IFU WITH NO ISSUES. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. THE LESION WAS CROSSED WITH MILD DIFFICULTY AND PRE-DILATED WITH A 4X60 EVERCROSS WITH NO ISSUES. THE 7X40 EVERCROSS WAS THEN INSERTED AND IT IS REPORTED THAT THE BALLOON BURST AT 7 ATMS ON FIRST INFLATION. THERE WERE REMOVAL DIFFICULTIES UPON DEFLATION AND THE CATHETER AND BALLOON WERE NOT ABLE TO BE REMOVED. NEGATIVE PRESSURE WAS THEN APPLIED TO BALLOON AGAIN WHERE IT WAS FOUND THAT BALLOON HAD RUPTURED. A BREAK/FRACTURE OCCURRED AT THE BALLOON CATHETER TIP. THE CATHETER WAS THEN ABLE TO BE REMOVED THROUGH 6 FR SHEATH, BU
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16982261 ↗MalfunctionNov 5, 2020Burst Container or Vessel; Burst Container or Vessel
PHYSICIAN INTENDED TO USE AN EVERCROSS PTA BALLOON DURING TREATMENT A PLAQUE LESION IN THE PATIENT¿S COMMON ILIAC ARTERY. SLIGHT VESSEL CALCIFICATION AND NO TORTUOSITY ARE REPORTED. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE PACKAGING. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. A NON-MEDTRONIC INFLATION DEVICE WAS USED FOR BALLOON INFLATION. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT. IT IS REPORTED THAT A BALLOON BURST OCCURRED AT AN INFLATION PRESSURE OF 10ATM. NO FRAGMENTS REMAIN IN THE PATIENT. A SECOND EVERCROSS BALLOON WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10792112 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page