EverCross™ AB35W07060080 (UDI 00643169974111)Recalls & reported problems
Ev3, Inc · PTA AB35W07060080 EVERCROSS 035 V08 · Class 2
All models of EverCross (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169974111), received between Sep 30, 2020 and Sep 30, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Detachment of Device or Device Component1
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2020Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component; Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component
PHYSICIAN INTENDED TO USE AN EVERCROSS PTA BALLOON WITH A 6FR NON-MEDTRONIC SHEATH DURING TREATMENT OF A 60MM CALCIFIED LESION IN THE PATIENT¿S PROXIMAL RIGHT COMMON ILIAC ARTERY. ARTERY DIAMETER REPORTED AS 7MM. MODERATE VESSEL CALCIFICATION AND TORTUOSITY ARE REPORTED. CALCIFICATION IS REPORTED THROUGHOUT THE LEG INCLUDED THE AFFECTED AREA. A NON-MEDTRONIC INFLATION DEVICE WAS USED FOR BALLOON INFLATION. EMBOLIC PROTECTION WAS NOT USED. AN ANGIOGRAM PERFORMED PRIOR TO BALLOON ANGIOPLASTY. THERE WAS NO DAMAGE NOTED TO PACKAGING PRIOR TO USE. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE PACKAGING. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS NOTED DURING ADVANCEMENT. THE BALLOON WAS PLACED THROUGH THE UP AND OVER SHEATH, FROM LEFT FEMORAL ACCESS. THE SHEATH WAS PULLED BACK, SO BALLOON WAS ACROSS THE ILIAC BIFURCATION. IT IS REPORTED A CIRCUMFERENTIAL/RADIAL BALLOON BURST OCCURRED AT A PRESSURE OF 12ATM ON FIRST INFLATION OF THE DEVICE. PHYSICIAN ATTEMPTED TO REMOVE BALLOON, BUT IT WAS STUCK ON THE CALCIFIED LESION. USING FORCE ON REMOVAL THE BALLOON CATHETER BROKE. THE MAIN SECT
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10605144 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page