EverCross™ AB35W07150135Recalls & reported problems

Ev3, Inc · PTA AB35W07150135 EVERCROSS 035 V08 · Class 2

All models of EverCross (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169974203), received between Mar 14, 2025 and Sep 30, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Burst Container or Vessel

    It was reported that during a procedure using an Evercross PTA balloon catheter, for the treatment of a plaque lesion in the mid section of the right superficial femoral artery (vessel diameter; 6.5 mm), the balloon experienced a pinhole type balloon burst (Balloon Burst AT, BELOW, or UNKNOWN Rated Burst Pressure) on the first inflation. The vessel was post- dilated. The device did not pass through a previously deployed stent. No resistance was encountered when advancing the device. The balloon was inflated with a (Cook Medical) inflation device and 50/50 contrast inflation fluid was used. No damage was noted to the packaging and no issues were noted when removing the device from the hoop/tray. The instructions for use were followed without issue. It is not known at what pressure the balloon burst. The balloon did not detach during the procedure. The device was safely removed from the patient. The procedure was completed using a new (Medtronic) device. The patient was reported as alive, with no injury. No injury or intervention was associated with this event. No patient symptoms or complications have been reported as a result of this event. It was reported that there was a 3 minute

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947116 ↗
  • MalfunctionMar 14, 2025Burst Container or Vessel; Burst Container or Vessel

    A PHYSICIAN WAS USING AN EVERCROSS PTA BALLOON FOR TREATMENT OF A SEVERELY CALCIFIED LESION IN THE MID RIGHT SUPERFICIAL FEMORAL ARTERY. A 6FR NON-MEDTRONIC SHEATH AND A 0.035 NON-MEDTRONIC GUIDEWIRE WERE USED. THE VESSEL WAS NOT PRE-DILATED. THE BALLOON PASSED THROUGH A PREVIOUSLY DEPLOYED EVERFLEX STENT. NO RESISTANCE WAS MET. A NON-MEDTRONIC INFLATION DEVICE WAS USED WITH 50/50 CONTRAST. IT WAS REPORTED DURING THE SECOND INFLATION THE BALLOON BURST LONGITUDINALLY. IT IS NOT KNOWN AT WHAT PRESSURE THE BALLOON BURST. THE BALLOON DID NOT DETACH AND IT WAS SAFELY REMOVED FROM THE PATIENT. ANOTHER EVERCROSS BALLOON WAS USED TO COMPLETE THE CASE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21604247 ↗

Other Catheter, percutaneous devices