NA 106A3 (UDI 00643169983892)Recalls & reported problems
Medtronic, Inc. · CONSOLE 106A3 GEN V ROHS PLX MKS · Class 3
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169983892), received between Oct 19, 2022 and Sep 5, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Infusion or Flow Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P020045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Infusion or Flow Problem1
- Protective Measures Problem1
- Temperature Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 5, 2024Infusion or Flow Problem; Protective Measures Problem; Infusion or Flow Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE SAFETY SYSTEM DETECTED A HIGH LEVEL OF REFRIGERANT FLOW AND STOPPED THE INJECTION. THE BALLOON CATHETER , COAXIAL UMBILICAL CABLE AND TANK WERE REPLACED WITHOUT RESOLVE. THE CASE WAS ABORTED THE PATIENT WAS UNDER GENERAL ANESTHESIA. A SERVICE WAS RECOMMENDED. DATA FILES WERE REVIEWED AND PRESSURE WAS NOT AS EXPECTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20144976 ↗InjuryAug 27, 2024Adverse Event Without Identified Device or Use Problem; Temperature Problem; Adverse Event Without Identified Device or Use Problem; Temperature Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THERE WAS NO LIMIT TO THE TEMPERATURE, AND IT DECREASED UNEXPECTEDLY. THE BALLOON CATHETER AND COAXIAL UMBILICAL CABLE WERE REPLACED WITHOUT RESOLVE. THE CASE WAS ABORTED. THE PATIENT WAS UNDER GENERAL ANESTHESIA. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20075546 ↗MalfunctionOct 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE REFRIGERANT DELIVERY PATH WAS OBSTRUCTED. THE PRESSURE AND FLOW WERE DECREASING. THE COAXIAL UMBILICAL CABLE WAS REPLACED WITHOUT RESOLVING THE ISSUES. THE HIGH PRESSURE REGULATOR WAS RISING. THE PROCEDURE WAS ABORTED UNDER GENERAL ANESTHESIA. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15632926 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020045. One submission often covers many models, so check the product and lot details in each record.
- Z-2777-2015 · Sep 1, 2015 · TerminatedMedtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.
- Z-1452-06 · Aug 4, 2006 · Terminated7F Freezor Cardiac Cryoablation catheter, REF # 207F1. CryoCath Technologies Inc.
Other Cardiac ablation percutaneous catheter devices
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- SMARTABLATE SAT001Biosense Webster Inc · Cardiac ablation percutaneous catheter185 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.